Abstract
-
Modular neck (MN) implants can restore the anatomy, especially in deformed hips such as sequelae of development dysplasia.
-
Early designs for MN implants had problems with neck fractures and adverse local tissue, so their use was restricted to limited indications.
-
Results of the latest generation of MN prostheses seem to demonstrate that these problems have been at least mitigated.
-
Given the results of the studies presented in this review, surgeons might consider MN total hip arthroplasty (THA) for a narrower patient selection when a complex reconstruction is required.
-
Long MN THA should be avoided in case of body mass index > 30, and should be used with extreme caution in association with high offset femoral necks with long or extra-long heads. Cr-Co necks should be abandoned, in favour of a titanium alloy connection.
-
Restoring the correct anatomic femoral offset remains a challenge in THA surgeries.
-
MN implants have been introduced to try to solve this problem. The MN design allows surgeons to choose the appropriate degree and length of the neck for desired stability and range of motion.
Cite this article: EFORT Open Rev 2021;6:751-758. DOI: 10.1302/2058-5241.6.200064
Introduction
Total hip arthroplasty (THA) is a beneficial surgical procedure that relieves pain, restores hip function, and improves quality of life in most patients with end-stage hip osteoarthritis or other hip diseases. 1 However, the procedure is not entirely failure-free and new designs are continually developing. 2,3 In the late 20th century, a new design has been introduced: the modular neck (MN) femoral stem. It is intended to give the surgeon the possibility to improve restoration of the joint biomechanics by adjusting the femoral version, the lower limb length, the neck-shaft angle and the femoral offset (FO), 4–7 defined as the distance from the centre of rotation of the femoral head to a line bisecting the longitudinal axis of the femur. 8–10
The theoretical benefits of FO restoration during total hip arthroplasty should be the reduction of bearing surface wear, implant loosening and dislocation rates. 4 The FO restoration is crucial to improve joint stability, functional outcome and implant longevity. 11,12 Therefore modularity, maximizing the options for a correct anatomic reconstruction, can be particularly useful in ‘difficult’ hips with advanced anatomical damage as in developmental dysplasia of the hip (DDH) to avoid complementary surgery such as osteotomies, or post-traumatic osteoarthritis, making the surgical treatment of a distorted hip easier, safer and more reproducible, reducing high morbidity and improving outcomes. 13,14
Despite the benefits of modularity, there are some significant disadvantages related to an increased risk of mechanical failure: dissociation at the neck–stem junction, 13 neck fractures, 15–17 fretting and corrosion at the neck–stem junction have been described. 18,19 Thus, the use of modular stems has been widely questioned and some models were even removed from the global market. 19,20 This review aims to describe the clinical outcomes and complications of MN primary total hip arthroplasty.
Materials and methods
The present study was conducted following the Preferential Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines. 21
Eligibility criteria
Inclusion criteria for studies in this systematic review were as follows: (1) published in English; (2) involved patients who underwent primary total hip arthroplasty surgery with modular prosthesis (3) reported aetiology or survival rate or bone stability or clinical score or complications or blood ion levels. Research was not limited in time; reviews, studies in vitro or biomechanical and cadaver experimentations were excluded.
Information sources and search
A systematic literature search of potentially eligible trials was conducted through online databases including PubMed-MEDLINE and the Cochrane Central Registry of Controlled Trials. The PubMed search included articles published online ahead of print. The utilized search strings were: ((((modular) AND ((“prosthesis implantation”[MeSH Terms]) OR ((“prosthesis”) AND “implantation”))) AND hip prosthesis[MeSH Terms]) AND ((hip prosthesis[MeSH Terms]) OR ((“hip”) AND “prosthesis”))) OR ((modular) AND ((“neck”) AND “prosthesis implantation”[MeSH Terms])). A manual search of related references and cited articles was also performed.
Screening and assessment of eligibility
Three reviewers (GV, MC and GO) independently screened the titles and abstracts of all studies for eligibility. Duplicate articles were manually excluded. All reviewers evaluated the full texts of all potentially eligible studies identified by title and abstract screening to determine final eligibility. All discrepancies were resolved by a consensus decision. The study selection process, carried out in accordance with PRISMA flow chart, is shown in Fig. 1.
Quality assessment of the studies
The quality of the studies was evaluated according to American Academy of Orthopedic Surgeons (AAOS) clinical practice guideline and systematic review methodology. 22 The following questions were used to evaluate the study quality of diagnostic study designs: (1) Was the patient spectrum representative of the patients who will receive the test in practice? (2) Were the selection criteria clearly described? (3) Was the execution of the index and reference tests described in sufficient detail to permit its replication? (4) Is the reference standard likely to correctly classify the target condition? (5) Are the index test results interpreted by an examiner without the knowledge of the reference tests results?
A study is considered as a high-quality study if it has < 1 flaw, as a moderate-quality study if it has ≥ 1 and < 2 flaws, as a low-quality study if it has ≥ 2 and < 3 flaws and as a very low-quality study if it has ≥ 3 flaws.
Results
Search results
The research we performed identified 1072 potentially eligible studies, and 9 additional records were found during manual searches of the reference lists. After removing 12 duplicates another 1035 studies were excluded based on their titles and abstracts. The remaining 34 articles were read in full, and 14 more articles were excluded. In the end 20 studies were included in this systematic review. 23–42 According to the AAOS clinical practice guideline and systematic review methodology, no studies have been rated as low or very low-quality studies.
Cohort characteristics
All the studies were analysed, and the extracted data are summarized in Table 1. A total of 5645 primary THA in 5518 patients were reported, the average follow-up was 71.37 months. Eighteen studies reported the primary diagnosis: primary osteoarthritis in 2745 (76.2%) cases, DDH in 526 (14.6%) cases, osteonecrosis in 64 (1.7%) cases, post-traumatic arthritis in 26 (0.75%) cases, rheumatoid arthritis in 17 (0.5%) cases and other diagnosis in 224 (6.2%) cases.
Details of included studies
Name | Year | N of patients | N of hips | Average follow-up (months) |
---|---|---|---|---|
Montalti et al 23 | 2018 | 80 | 80 | 181.2 |
Pour et al 24 | 2016 | 244 | 277 | 50.0 |
Blakey et al 33 | 2009 | 288 | 316 | 86.4 |
Loubignac et al 34 | 2005 | 68 | 76 | 80.4 |
Gofton et al 35 | 2017 | 809 | 809 | 68.4 |
Haversath et al 36 | 2017 | 93 | 93 | 24.0 |
Mikkelsen et al 37 | 2017 | 33 | 33 | 27.6 |
Collet et al 38 | 2017 | 72 | 74 | 112.8 |
Vanbiervliet et al 39 | 2017 | 86 | 95 | 78.0 |
Bernstein et al 40 | 2016 | 63 | 73 | 50.4 |
Laurençon et al 41 | 2016 | 40 | 40 | 12.0 |
Nawabi et al 42 | 2016 | 199 | 199 | 19.3 |
Ollivier et al 25 | 2015 | 170 | 170 | 71.0 |
Silverton et al 26 | 2014 | 152 | 152 | 54.0 |
Molloy et al 27 | 2014 | 16 | 15 | 42.3 |
Duwelius et al 28 | 2014 | 459 | 459 | 28.8 |
Cossetto et al 29 | 2012 | 138 | 162 | 93.6 |
Traina et al 30 | 2009 | 2131 | 2131 | 108.0 |
Pelayo-de-Tomás et al 31 | 2018 | 317 | 317 | 73.2 |
Sakai et al 32 | 2010 | 60 | 74 | 174.0 |
Total | 5518 | 5645 | 71.4 |
Surgical approach
Seventeen studies reported the surgical approach for a total of 3870 procedures. The most frequently used approach was the posterolateral in 1572 (40%) cases, a posterior approach was used in 895 (23%) cases, an antero-lateral approach in 622 (16%) cases, a direct anterior approach in 506 (13%) cases and a lateral approach in the last 275 (7%) cases.
Survival rate and bone stability
Thirteen studies reported a Kaplan-Meier (KM) analysis of survival rate. Montalti et al 23 with the longest follow-up registered (15 years) presented a survival rate of 90.5%, with only two cases of aseptic loosening out of 80 implants. Also Collet et al 38 with a rate of 94.2%, with any revision or reoperation as endpoint, registered a good survival of the stems used. Cossetto et al 29 reported 10-year KM survival of 96% for the MBA stem; Blakey et al 33 reported five-year KM survival of 97.5% for the ANCA-Fit stem; and Silverton et al 26 reported eight-year KM survival of 89.4% for the uncemented modular Profemur® Z stem. Conversely Bernstein et al 40 in 2016 described an 86% clinical failure rate (63 of the 73 hips) at a mean follow-up of 4.2 ± 0.6 years (range, 3.0 to 5.5 years). The radiographic analysis was reported in 11 studies showing a mean result of 99.56% (97.5–100%). All data extracted are summarized in Table 2.
Survival rate and bone stability
Name | Year | Survival rate (%) | Bone stability (%) |
---|---|---|---|
Montalti et al 23 | 2018 | 90.5 | 97.5 |
Pour et al 24 | 2016 | 85.0 | NR |
Blakley et al 33 | 2009 | 97.5 | 98.0 |
Loubignac et al 34 | 2005 | 90.0 | NR |
Collet et al 38 | 2017 | 94.2 | 100.0 |
Vanbiervliet et al 39 | 2017 | 100.0 | 100.0 |
Bernstein et al 40 | 2016 | 0.0 | NR |
Nawabi et al 42 | 2016 | 69.3 | NR |
Ollivier et al 25 | 2015 | 98.9 | 100.0 |
Silverton et al 26 | 2014 | 89.4 | NR |
Cossetto et al 29 | 2012 | 99.0 | 100.0 |
Traina et al 30 | 2009 | 96.8 | NR |
Pelayo et al-de-Tomás 31 | 2018 | 95.0 | 99.0 |
Note. NR, not reported.
Clinical results
Fourteen studies reported clinical outcomes using the Harris Hip Score (HHS) and the Merle d’Aubigne Postel Score (MAPS). The HHS results are summarized in Table 3. We estimated a mean preoperative score of 45.4 and a mean postoperative score of 88.8 points (80. 7–98.6). Collet et al, 38 in their series, registered a mean postoperative HHS of 93.6±8.2 at a mean follow-up of 9.4 years, which was greater than the results reported by several other authors. 26,30,33 The MAPS score summarized in Table 3 improved from an average preoperative score of 10 to a postoperative mean score of 17.5 (15–18).
Harris Hip Score (HHS) and Merle d’Aubigne score
Name | Year | Preoperative Harris Hip Score | Postoperative Harris Hip Score | Preopearative Merle d’Aubigne score | Postoperative Merle d’Aubigne score |
---|---|---|---|---|---|
Montalti et al 23 | 2018 | 35.7 | 82.8 | 9.3 | 15.1 |
Loubignac et al 34 | 2005 | NR | 91.3 | 16.4 | 18.0 |
Haversath et al 36 | 2017 | 50.9 | 93.8 | – | – |
Collet et al 38 | 2017 | NR | 93.6 | – | – |
Vanbiervliet et al 39 | 2017 | 31.0 | 90.0 | – | – |
Nawabi et al 42 | 2016 | 50.7 | 81.8 | – | – |
Ollivier et al 25 | 2015 | 50.4 | 84.5 | – | – |
Molloy et al 27 | 2014 | NR | 80.7 | – | – |
Duwelius et al 28 | 2014 | 52.0 | 91.0 | – | – |
Sakai et al 32 | 2010 | 44.6 | 98.6 | – | – |
Cossetto et al 29 | 2012 | – | – | 7.1 | 16.4 |
Average | 45 | 88.8 | 10.0 | 17.5 |
Ion serum levels
Five studies 24,26,31,41,42 reported cobalt and chromium ion serum levels in ng/ml, that are summarized in Table 4. Only one paper 41 evaluated systemic metal ion levels in a homogenous series of patients after implantation of a mixed-alloy modular neck hip prosthesis (Symbios SPS, Ti stems coupled to Co-Cr necks) compared to those of two control groups: a group of patients having received the non-modular neck version of the same stem (Ti6Al4V) and a group of patients without any metal implant.
CoCr ion serum levels
Name | Year | N of patients | CoCr ion levels (ng/ml) | Material of the heads | Material of the necks |
---|---|---|---|---|---|
Pour et al 24 | 2016 | 244 | Co 5.5 Cr 3.16 |
62 ceramic 182 Co-Cr alloy |
Co-Cr |
Laurençon et al 41 | 2016 | 40 | Co 1.28±2.32 Cr 1.12±0.52 |
Ceramic | Co-Cr-molybdenum alloy |
Nawabi et al 42 | 2016 | 199 | Co 6.1 Cr 1.3 |
Co-Cr alloy | Co-Cr alloy |
Silverton et al 26 | 2014 | 152 | Co 2.4 Cr 1.7 |
30 ceramic 122 Co-Cr |
Titanium alloy |
Pelayo-de-Tomás et al 31 | 2018 | 317 | Co < 2.2 Cr < 2.2 |
297 ceramic 20 metal |
79 titanium 238 Cr-Co |
The higher cobalt ion serum level was registered by Nawabi et al 42 where serum metal ion levels were obtained in 144 of the 195 patients. The mean serum cobalt was 6.1 ng/mL (range, 0.5 to 7.7 ng/mL), and the mean serum chromium was 1.3 ng/mL (range, 0.2 to 5.3 ng/mL) higher than the normal (serum cobalt < 1 ng/mL, serum chromium < 5 ng/mL). Patients who had undergone revision had significantly higher serum cobalt ion levels (mean, 8.6 ng/mL).
Also Pour et al, 24 in their series, described greater serum levels of Cr and Ti considering all hips requiring aseptic revision surgery for any reason (n = 43 of 277).
Complications
Thirteen studies reported complications for a total of 202 episodes in 3106 THA: adverse local tissue reaction (ALTR) in 85 (38%) cases, aseptic loosening of the stem or acetabular component in 28 (13.86%) cases, acute or chronic dislocations in 31 (15.35%) cases, periprosthetic fractures in 21 (1.4%) cases, deep infections in 10 (4.95%) cases, three studies reported neck fracture in 12 (5.94%) cases, pain and impingement respectively in two (0.99%) and two (0.99%) cases, ceramic head fractures in one (0.5%) case and one (0.5%) case of varus positioning of the stem. Other complications were reported in four (1.98%) cases. All complications have been summarized in Table 5.
Complications
Name | Year | N of hips | Complications |
---|---|---|---|
Montalti et al 23 | 2018 | 80 | Aseptic loosening 2 (2.5%) Dislocations 2 (2.5%) Periprosthetic fracture 1 (1.25%) |
Pour et al 24 | 2016 | 277 | Aseptic loosening 15 (5%) Neck fractures 7 (2.5%) Periprosthetic fractures 4 (1.5%) ALTR 4 (1.5%) |
Blakley et al 33 | 2009 | 316 | Dislocations 5 (1.5%) Aseptic loosening 4 (1.2%) |
Loubignac et al 34 | 2005 | 76 | Dislocations 3 (4%) Periprosthetic fractures 1 (1.3%) Aseptic loosening 1 (1.3%) Other 3 (4%) |
Gofton et al 35 | 2017 | 809 | Dislocations 9 (1.1%) Neck fractures 4 (0.05%) |
Nawabi et al 42 | 2016 | 199 | ALTR 73 (36%) Periprosthetic fractures 5 (2.5%) Pain 2 (1%) |
Ollivier et al 25 | 2015 | 170 | Deep infections 3 (1.7%) Periprosthetic fractures 2 (1.1%) Impingement 2 (1.1%) |
Silverton et al 26 | 2014 | 152 | 2 taper corrosion 1 cup malposition 1 ALTR 1 neck fracture (0.66%) |
Molloy et al 27 | 2014 | 15 | ATLR 7 (46.6%) |
Duwelius et al 28 | 2014 | 459 | Dislocations 2 (0.4%) Deep infections 2 (0.4%) Intraoperative fracture 1 (0.2%) Other 1 (0.2%) |
Cossetto et al 29 | 2012 | 162 | Periprosthetic fractures 5 (3%) Dislocations 3 (1.8%) Deep infections 2 (1.2%) Aseptic loosening 2 (1.2%) |
Pelayo-de-Tomás et al 31 | 2018 | 317 | Dislocations 7 (2.2%) Periprosthetic fractures 3 (0.9%) Deep infections 3 (0.9%) Intraoperative fracture 1 (0.3%) |
Sakai et al 32 | 2010 | 74 | Aseptic loosening 6 (8%) |
Note. ALTR, adverse local tissue reaction.
Discussion
THA with MN stem appears to be a useful procedure for the treatment of hip osteoarthritis or other hip diseases. 1 There are remarkable improvements in functional outcomes in the form of HHS, MEPS and range of motion in the papers reviewed. MN stems can play a major role in case of anatomical abnormalities of the proximal femur. 43 The average improvement in HHS was 43.76 points (range 28–59 points). Nawabi et al 42 reported the lowest postoperative average improvement in HHS (31.1 points).
In our review, the crude overall complication rate (202 in 3106 cases) was 6.5%. This percentage was slightly higher than that reported in literature with the use of a monobloc femoral stem. 44 As described in the literature, comparable complication rates of deep infections 0.32% (10 in 3106 cases), periprosthetic fracture 0.68% (21 of 3106 cases) were registered by the studies of this review.
In the Australian Orthopaedic Association National Joint Replacement Registry of 2020, 10,235 primary procedures were recorded using femoral stems with modular necks. The cumulative percent revision at 15 years was 12.5% for modular stems; double compared to 7.7% for fixed neck stems. 45 ALTR was the most frequent complication reported (85 cases, 2.73%). The majority of these cases (73 of 85) were recorded by Nawabi et al, 42 who conclude that the corrosion at the neck–stem junction in total hip arthroplasty could be a cause of ALTR. The authors confirm that their findings may apply to a particular specific modular design and suggest that surveillance utilizing metal ion levels and MRI (magnetic resonance imaging) may be indicated for follow-up patients with modular neck hip prostheses.
It is described that Cr-Co neck corrosion determines the release of metal ions that results in ALTR. This has resulted in the voluntary recall of many MN total hip arthroplasty designs. Only five of the 23 papers included in this review reported the serum metal ion levels data (Table 4) 24,33,34,36,41 with levels higher than the normal (serum cobalt < 1 ng/mL, serum chromium < 5 ng/mL).
In the Australian Orthopaedic Association National Joint Replacement Registry of 2020, femoral stem with modular necks still have almost twice the rate of revision compared to fixed neck stems. The titanium/cobalt chromium combination has an extremely high rate of revision compared to the titanium/titanium combination. At 15 years, the cumulative incidence of metal-related pathology was 3.9% for titanium/cobalt chromium compared to 0.1% for titanium/titanium. 45
The overall dislocation rate recorded as 0.93% (29 of 3106 cases) is slightly lower in comparison to that described in the literature (2–4%). 46–48 Only one article, by Gofton et al, 35 correlated the surgical approach to the dislocation rate. In their study there was a significant difference between dislocation rates based on approach (p < 0.05). There were no dislocations with a lateral approach, one dislocation with the anterior approach and eight with the posterior approach (2.33%, 8/343).
As described by different authors 5,10,38 the modularity was effective in restoring offset and leg length, and these confer stability with a good mid to long-term follow-up. Traina et al 49 reported that MN prostheses allow the restoration of femoral offset, abductors moment arm, leg length and joint kinematics especially in critical cases of THA, such as high-grade dysplasia. Restoring the proper offset appears to determinate an appropriate tension of the abductor muscles, which implies a better functional outcome and a better primary stability of the implant with less early migration. 10 These results are in contrast with those reported by Carothers et al, who found significant differences of more than 4 mm in neck length and more than 2 mm in offset in only 15% of cases in their study. 48
These conclusions suggest that use of modular heads results in head centre positions also achievable with non-modular stems with a correct preoperative planning in most cases. 50 Also, different studies did not demonstrate an effective impact of modularity on dislocation. In a retrospective series of 809 THAs using the Profemur TM stem (Wright®) with modular neck, Gofton et al 35 reported 2.3% dislocation on a posterior approach, compared to 0.3% on anterior and 0% on lateral approaches, whereas modularity was most often used in posterior approaches. Comparing two consecutive series of a single surgeon, with 284 standard and 594 modular neck THAs, Duwelius et al 5 found no difference in Harris or SF12 scores at a mean 2.4 years follow-up.
Fracture of the femoral component is a rare complication in total hip arthroplasty. In our review we registered 12 cases of neck fracture with a rate of 0.38%. 26,28,37 Traina et al 49 confirm that the failure of modular necks is related to neck offset, stem size (the larger the stem, the higher the risk) and body weight. The authors suggest that in the presence of a high BMI, young male patient, a caput-collum-diaphyseal (CCD) angle < 135°, and a high functional demand the risk of failure is significantly increased, and the use of modular prostheses must be evaluated case by case. Konan et al, 51 in their review, also confirm that the incidence of this complication is low, but the risk increases in patients with a high BMI, a high level of activity, a small medullary canal (hence small stem diameters) and those with severe bone loss and poor proximal bony support for the stem, especially medially.
In a systematic review of the literature and of English-language registries, Mihalko et al showed that stems with modular necks had significantly poorer 10-year survival than monobloc stems with cumulative revision rates between 4% and 9%, 50 particularly when modular neck stems were paired with metal-on-metal articulating bearing, which was reported to increase revision rates even further. 45 The authors conclude that it remains to be established whether this is the result of implant taper mismatch in the assembly of the added taper junction, implant material or design, or surgical technical errors.
A 2020 scoping review by Mertl et al 52 concludes that modular neck stems have to imperatively be reserved for difficult cases of dysplasia or severely deformed femur, due to the risk of fracture. They are to be avoided in case of obese patients (BMI > 30), Cr-Co necks should be abandoned, in favour of a Ti6Al4V connection. Cr-Co/Ti connections are associated with serious corrosion, leading to severe adverse local tissue reaction.
Conclusion
Despite the concerns among most of surgeons in several countries, the results of our review suggest that MN stem prosthesis in primary THA is a useful treatment for hip osteoarthrosis in case of DDH or severely deformed femur. The majority of the published studies report clinical outcomes after THA with MN prosthesis comparable to those reported with the use of monobloc stems. The modularity of the necks in narrower selected cases allows the restoration of femoral offset, abductors moment arm and leg length that could result in better functioning of the joint and better primary stability of the implant. This should result in good mid to long-term follow-up outcomes. The new generations of implant and the use of the Ti6Al4V neck reduce the risk of wear and adverse local tissue reaction that remains the most common complication registered in the analysed studies.
It is a fact that the risk of wear and neck-taper fractures is higher in obese male patients with a BMI > 30 and high functional demand, and the use of modular stems in these patients is to be avoided. The use of long and extra-long necks must be carefully evaluated as well. Future high-quality prospective studies with longer follow-up are necessary to confirm the supposed advantages of the use of MN stem in difficult primary THA procedures and to highlight all disadvantages with respect to NMN THA.
Open access
This article is distributed under the terms of the Creative Commons Attribution-Non Commercial 4.0 International (CC BY-NC 4.0) licence (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed.
LZ reports consultancy fees paid by Medacta for product development; fees paid for lectures by Zimmer Biomet, Stryker, LimaCorporate and Smith & Nephew, and research fees paid to his institution by LimaCorporate, DePuy and Medacta – all for relevant financial activities outside the submitted work. All other authors declare no conflicts of interest relevant to this work.
No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.
This article is distributed under the terms of the Creative Commons Attribution-Non Commercial 4.0 International (CC BY-NC 4.0) licence (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed.
References
- 1.↑
Learmonth ID , Young C , Rorabeck C . The operation of the century: total hip replacement. Lancet 2007; 370:1508–1519 .
- 2.↑
Bozic KJ , Kurtz SM , Lau E , Ong K , Vail TP , Berry DJ . The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg [Am] 2009; 91-A:128–133 .
- 3.↑
Sanchez-Sotelo J , Berry DJ . Epidemiology of instability after total hip replacement. Orthop Clin North Am 2001; 32:543–552, vii.
- 4.↑
Traina F , De Fine M , Biondi F , Tassinari E , Galvani A , Toni A . The influence of the centre of rotation on implant survival using a modular stem hip prosthesis. Int Orthop 2009; 33:1513–1518 .
- 5.↑
Duwelius PJ , Hartzband MA & Burkhart R et al. Clinical results of a modular neck hip system: hitting the ‘bull’s-eye’ more accurately. Am J Orthop 2010; 39:2–6 .
- 6.
Asayama I , Chamnongkich S , Simpson KJ , Kinsey TL , Mahoney OM . Reconstructed hip joint position and abductor muscle strength after total hip arthroplasty. J Arthroplasty 2005; 20:414–420 .
- 7.↑
McGrory BJ , Morrey BF , Cahalan TD , An KN , Cabanela ME . Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty. J Bone Joint Surg [Br] 1995; 77-B:865–869 .
- 8.↑
Lecerf G , Fessy MH & Philippot R et al. Femoral offset: anatomical concept, definition, assessment, implications for preoperative templating and hip arthroplasty. Orthop Traumatol Surg Res 2009; 95:210–219 .
- 9.
Charles MN , Bourne RB , Davey JR , Greenwald AS , Morrey BF , Rorabeck CH . Soft-tissue balancing of the hip: the role of femoral offset restoration. Instr Course Lect 2005; 54:131–141 .
- 10.↑
Vicenti G , Solarino G & Spinarelli A et al. Restoring the femoral offset prevent early migration of the stem in total hip arthroplasty: an EBRA-FCA study. J Biol Regul Homeost Agents 2016; 30:S207–S212 .
- 11.↑
Husby VS , Bjørgen S , Hoff J , Helgerud J , Benum P , Husby OS . Unilateral vs. bilateral total hip arthroplasty: the influence of medial femoral head offset and effects on strength and aerobic endurance capacity. Hip Int 2010; 20:204–214 .
- 12.↑
Chamnongkich S , Asayama I , Kinsey TL , Mahoney OM , Simpson KJ . Difference in hip prosthesis femoral offset affects hip abductor strength and gait characteristics during obstacle crossing. Orthop Clin North Am 2012; 43:e48–e58 .
- 13.↑
Kouzelis A , Georgiou CS , Megas P . Dissociation of modular total hip arthroplasty at the neck-stem interface without dislocation. J Orthop Traumatol 2012; 13:221–224 .
- 14.↑
Mencière M-L , Amouyel T , Taviaux J , Bayle M , Laterza C , Mertl P . Fracture of the cobalt-chromium modular femoral neck component in total hip arthroplasty. Orthop Traumatol Surg Res 2014; 100:565–568 .
- 15.↑
Sotereanos NG , Sauber TJ , Tupis TT . Modular femoral neck fracture after primary total hip arthroplasty. J Arthroplasty 2013; 28:196.e7–196.e9 .
- 16.
Kop AM , Swarts E . Corrosion of a hip stem with a modular neck taper junction: a retrieval study of 16 cases. J Arthroplasty 2009; 24:1019–1023 .
- 17.↑
Ellman MB , Levine BR . Fracture of the modular femoral neck component in total hip arthroplasty. J Arthroplasty 2013; 28:196.e1–196.e5 .
- 18.↑
Gilbert JL , Buckley CA , Jacobs JJ . In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations: the effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res 1993; 27:1533–1544 .
- 19.↑
Walsh CP , Hubbard JC , Nessler JP , Markel DC . Revision of recalled modular neck rejuvenate and ABG femoral implants. J Arthroplasty 2015; 30:822–826 .
- 20.↑
Meftah M , Haleem AM , Burn MB , Smith KM , Incavo SJ . Early corrosion-related failure of the rejuvenate modular total hip replacement. J Bone Joint Surg [Am] 2014; 96-A:481–487 .
- 21.↑
Moher D , Liberati A , Tetzlaff J , Altman DG ; PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. PLoS Med 2009; 6:e1000097 .
- 22.↑
Surgeons AAOS clinical practice guideline and systematic review methodology. OrthoGuidelines. https://www.aaos.org/uploadedFiles/PreProduction/Quality/Guidelines_and_Reviews/guidelines/GuidelineandSystematicReviewProcesses_v2.0_Final.pdf (date last accessed 22 September 2019).
- 23.↑
Montalti M , Castagnini F , Giardina F , Tassinari E , Biondi F , Toni A . Cementless total hip arthroplasty in Crowe III and IV dysplasia: high hip center and modular necks. J Arthroplasty 2018; 33:1813–1819 .
- 24.↑
Pour AE , Borden R , Murayama T , Groll-Brown M , Blaha JD . High risk of failure with bimodular femoral components in THA. Clin Orthop Relat Res 2016; 474:146–153 .
- 25.↑
Ollivier M , Parratte S , Galland A , Lunebourg A , Flecher X , Argenson J-N . Titanium-titanium modular neck for primary THA: result of a prospective series of 170 cemented THA with a minimum follow-up of 5 years. Orthop Traumatol Surg Res 2015; 101:137–142 .
- 26.↑
Silverton CD , Jacobs JJ , Devitt JW , Cooper HJ . Midterm results of a femoral stem with a modular neck design: clinical outcomes and metal ion analysis. J Arthroplasty 2014; 29:1768–1773 .
- 27.↑
Molloy DO , Munir S , Jack CM , Cross MB , Walter WL , Walter WK Sr . Fretting and corrosion in modular-neck total hip arthroplasty femoral stems. J Bone Joint Surg [Am] 2014; 96-A:488–493 .
- 28.↑
Duwelius PJ , Burkhart B & Carnahan C et al. Modular versus nonmodular neck femoral implants in primary total hip arthroplasty: which is better? Clin Orthop Relat Res 2014; 472:1240–1245 .
- 29.↑
Cossetto DJ , Goudar A . Mid-term outcome of a modular, cementless, proximally hydroxyapatite-coated, anatomic femoral stem. J Orthop Surg (Hong Kong) 2012; 20:322–326 .
- 30.↑
Traina F , De Clerico M , Biondi F , Pilla F , Tassinari E , Toni A . Sex differences in hip morphology: is stem modularity effective for total hip replacement? J Bone Joint Surg [Am] 2009; 91-A:121–128 .
- 31.↑
Pelayo-de-Tomás JM , Rodrigo-Pérez JL , Novoa-Parra CD , Lizaur-Utrilla A , Morales-Suárez-Varela M , Blas-Dobón JA . Cementless modular neck stems: are they a safe option in primary total hip arthroplasty? Eur J Orthop Surg Traumatol 2018; 28:463–469 .
- 32.↑
Sakai T , Ohzono K , Nishii T , Miki H , Takao M , Sugano N . A modular femoral neck and head system works well in cementless total hip replacement for patients with developmental dysplasia of the hip. J Bone Joint Surg [Br] 2010; 92-B:770–776 .
- 33.↑
Blakey CM , Eswaramoorthy VK , Hamilton LC , Biant LC , Field RE . Mid-term results of the modular ANCA-Fit femoral component in total hip replacement. J Bone Joint Surg [Br] 2009; 91-B:1561–1565 .
- 34.↑
Loubignac F , Rebouillat J . Advantages of the bimodular neck for THR: results of the first 100 Hélianthe femoral stems to at least 5-year follow-up. Eur J Orthop Surg Traumatol 2005; 15:90–100 .
- 35.↑
Gofton WT , Illical EM , Feibel RJ , Kim PR , Beaulé PE . A single-center experience with a titanium modular neck total hip arthroplasty. J Arthroplasty 2017; 32:2450–2456 .
- 36.↑
Haversath M , Wendelborn C , Jäger M , Schmidt B , Kowalczyk W , Landgraeber S . Geometric analysis and clinical outcome of two cemented stems for primary total hip replacement with and without modular necks. Arch Orthop Trauma Surg 2017; 137:1571–1578 .
- 37.↑
Mikkelsen RT , Fløjstrup M & Klem N et al. Modular neck vs nonmodular femoral stems in total hip arthroplasty: clinical outcome, metal ion levels, and radiologic findings. J Arthroplasty 2017; 32:2774–2778 .
- 38.↑
Collet T , Atanasiu J-P & de Cussac J-B et al. Midterm outcomes of titanium modular femoral necks in total hip arthroplasty. Ann Transl Med 2017; 5:395 .
- 39.↑
Vanbiervliet J , Somers JFA . Excellent mid-term results of a new polished tapered modular cemented stem: a study of 113 hip replacements with minimum 5-year follow-up. Hip Int 2017; 27:455–459 .
- 40.↑
Bernstein DT , Meftah M , Paranilam J , Incavo SJ . Eighty-six percent failure rate of a modular-neck femoral stem design at 3 to 5 years: lessons learned. J Bone Joint Surg [Am] 2016; 98-A:e49 .
- 41.↑
Laurençon J , Augsburger M , Faouzi M , Becce F , Hassani H , Rüdiger HA . Systemic metal ion levels in patients with modular-neck stems: a prospective cohort study. J Arthroplasty 2016; 31:1750–1755 .
- 42.↑
Nawabi DH , Do HT & Ruel A et al. Comprehensive analysis of a recalled modular total hip system and recommendations for management. J Bone Joint Surg [Am] 2016; 98-A:40–47 .
- 43.↑
Zagra L . Advances in hip arthroplasty surgery: what is justified? EFORT Open Rev 2017; 2:171–178 .
- 44.↑
Loppini M , Grappiolo G . Uncemented short stems in primary total hip arthroplasty: the state of the art. EFORT Open Rev 2018; 3:149–159 .
- 45.↑
Australian Orthopaedic Association National Joint Replacement Registry. Annual report 2019. Adelaide, 2020. https://aoanjrr.sahmri.com/annual-reports-2019 (Date last accessed 28 August 2021).
- 46.↑
Lübbeke A , Suvà D , Perneger T , Hoffmeyer P . Influence of preoperative patient education on the risk of dislocation after primary total hip arthroplasty. Arthritis Rheum 2009; 61:552–558 .
- 47.
Meek RMD , Allan DB , McPhillips G , Kerr L , Howie CR . Late dislocation after total hip arthroplasty. Clin Med Res 2008; 6:17–23 .
- 48.↑
Carothers JT , Archibeck MJ , Tripuraneni KR . Modular versus nonmodular femoral necks for primary total hip arthroplasty. Am J Orthop 2015; 44:411–414 .
- 49.↑
Traina F , De Fine M , Tassinari E , Sudanese A , Calderoni PP , Toni A . Modular neck prostheses in DDH patients: 11-year results. J Orthop Sci 2011; 16:14–20 .
- 50.↑
Mihalko WM , Wimmer MA , Pacione CA , Laurent MP , Murphy RF , Rider C . How have alternative bearings and modularity affected revision rates in total hip arthroplasty? Clin Orthop Relat Res 2014; 472:3747–3758 .
- 51.↑
Konan S , Garbuz DS , Masri BA , Duncan CP . Modular tapered titanium stems in revision arthroplasty of the hip: the risk and causes of stem fracture. J Bone Joint Surg [Br] 2016; 98-B:S50–S53 .