Short-term success of proximal bone stock preservation in short hip stems: a systematic review of the literature

in EFORT Open Reviews
Authors:
Sheryl de Waard Specialized Centre of Othopedic Research & Education (SCORE) and Xpert Orthopedie, Amsterdam, Netherlands
Academic Medical Centre (AMC), Amsterdam, Netherlands

Search for other papers by Sheryl de Waard in
Current site
Google Scholar
PubMed
Close
,
Jacqueline van der Vis Specialized Centre of Othopedic Research & Education (SCORE) and Xpert Orthopedie, Amsterdam, Netherlands

Search for other papers by Jacqueline van der Vis in
Current site
Google Scholar
PubMed
Close
,
Pascale A.H.T. Venema Specialized Centre of Othopedic Research & Education (SCORE) and Xpert Orthopedie, Amsterdam, Netherlands

Search for other papers by Pascale A.H.T. Venema in
Current site
Google Scholar
PubMed
Close
,
Inger N. Sierevelt Specialized Centre of Othopedic Research & Education (SCORE) and Xpert Orthopedie, Amsterdam, Netherlands

Search for other papers by Inger N. Sierevelt in
Current site
Google Scholar
PubMed
Close
,
Gino M.M.J. Kerkhoffs Academic Medical Centre (AMC), Amsterdam, Netherlands

Search for other papers by Gino M.M.J. Kerkhoffs in
Current site
Google Scholar
PubMed
Close
, and
Daniël Haverkamp Specialized Centre of Othopedic Research & Education (SCORE) and Xpert Orthopedie, Amsterdam, Netherlands

Search for other papers by Daniël Haverkamp in
Current site
Google Scholar
PubMed
Close

Sheryl de Waard, Orthopaedic Surgery, SCORE, Laarderhoogtweg 12, 1101 AE Amsterdam, Netherlands. Email: s.dewaard21@gmail.com
Open access

  • Total hip arthroplasty is performed more frequently in younger patients nowadays, making long-term bone stock preservation an important topic. A mechanism for late implant failure is periprosthetic bone loss, caused by stress shielding around the hip stem due to different load distribution. Short stems are designed to keep the physical loading in the proximal part of the femur to reduce stress shielding. The aim of this review is to give more insight into how short and anatomic stems behave and whether they succeed in preservation of proximal bone stock.

  • A systematic literature search was performed to find all published studies on bone mineral density in short and anatomic hip stems. Results on periprosthetic femoral bone mineral density, measured with dual-energy X-ray absorptiometry (DEXA), were compiled and analysed per Gruen zone in percentual change.

  • A total of 29 studies were included. In short stems, Gruen 1 showed bone loss of 5% after one year (n = 855) and 5% after two years (n = 266). Gruen 7 showed bone loss of 10% after one year and –11% after two years. In anatomic stems, Gruen 1 showed bone loss of 8% after one year (n = 731) and 11% after two years (n = 227). Gruen 7 showed bone loss of 14% after one year and 15% after two years.

  • Short stems are capable of preserving proximal bone stock and have slightly less proximal bone loss in the first years, compared to anatomic stems.

Cite this article: EFORT Open Rev 2021;6:1040-1051. DOI: 10.1302/2058-5241.6.210030

Abstract

  • Total hip arthroplasty is performed more frequently in younger patients nowadays, making long-term bone stock preservation an important topic. A mechanism for late implant failure is periprosthetic bone loss, caused by stress shielding around the hip stem due to different load distribution. Short stems are designed to keep the physical loading in the proximal part of the femur to reduce stress shielding. The aim of this review is to give more insight into how short and anatomic stems behave and whether they succeed in preservation of proximal bone stock.

  • A systematic literature search was performed to find all published studies on bone mineral density in short and anatomic hip stems. Results on periprosthetic femoral bone mineral density, measured with dual-energy X-ray absorptiometry (DEXA), were compiled and analysed per Gruen zone in percentual change.

  • A total of 29 studies were included. In short stems, Gruen 1 showed bone loss of 5% after one year (n = 855) and 5% after two years (n = 266). Gruen 7 showed bone loss of 10% after one year and –11% after two years. In anatomic stems, Gruen 1 showed bone loss of 8% after one year (n = 731) and 11% after two years (n = 227). Gruen 7 showed bone loss of 14% after one year and 15% after two years.

  • Short stems are capable of preserving proximal bone stock and have slightly less proximal bone loss in the first years, compared to anatomic stems.

Cite this article: EFORT Open Rev 2021;6:1040-1051. DOI: 10.1302/2058-5241.6.210030

Introduction

Nowadays, there is a higher incidence of younger and more active patients undergoing total hip arthroplasty (THA). In the Netherlands, between 17% and 19% of the patients receiving total hip arthroplasty are younger than 60 years.1,2 Their biomechanical forces on the hip implant are higher compared to the more elderly with THA, increasing the long-term failure rate of hip implants.36 Failure of hip implants can be caused by periprosthetic bone loss, leading to aseptic loosening or periprosthetic fractures. Stress shielding is one of the reasons for periprosthetic bone loss in cementless total hip arthroplasty,7,8 as bone resorption occurs through changed physical loading in the femur.9,10 Loss of bone stock can compromise the stability of hip stems,11 emphasizing the importance of bone stock preservation around hip stems. Physical loading in most conventional hip stems is transferred from the proximal region of the femur to the femur shaft, decreasing bone mass around the greater trochanter and calcar region.12,13

Other types of implants, such as short and anatomical stem designs, are based on a more proximal anchoring principle. Short stems in total hip arthroplasty were designed to restrict the physical loading to the proximal part of the femur, thereby reducing periprosthetic bone loss.14 However, there are a lot of designs with different implant characteristics. A clear definition of a short stem still lacks. A classification based on the osteotomy level is currently used,15 but neglects the anchoring principle of the stem. Hip stems that are classified as short do not necessarily yield better proximal bone stock preservation compared to anatomical or conventional hip stems.

Dual-energy X-ray absorptiometry (DEXA) is a reliable and widely accepted tool to determine bone mineral density (BMD).1618 Measuring BMD around the total hip prosthesis provides information on the redistribution of the biomechanical forces in the femur as a result of implanted hip stems.19 A review by Knutsen et al13 gave insight into stress shielding around cemented and uncemented conventional hip stems, as all available data on bone remodelling were pooled. For uncemented conventional stems, the area of the greater trochanter and calcar region are most prone to bone loss. Results of proximal bone stock preservation for different types of short and anatomic stems have been published in recent years, with diverse outcomes, but there are no global data on their outcome. In this review, short-term results on BMD surrounding short stems were assessed in comparison with anatomic stems.

Methods

Inclusion criteria

Studies

A systematic literature search was performed to obtain all published data on bone mineral density (BMD) in total hip arthroplasty. Inclusion was limited to studies on BMD measurements surrounding proximal anchored femoral hip stems, with a follow-up period of at least one year after surgery. Studies on hip resurfacing and conventional hip stems were excluded. Prospective and retrospective cohorts, as well as comparative studies, were included.

Outcome measurement

Primary outcome was BMD loss in Gruen zone 1 (calcar region) and Gruen zone 7 (greater trochanter),20 representing proximal loading zones, as these zones can be compared per different stem design. The outcome was described as the relative percentual change in BMD per follow-up moment relative to baseline. Baseline value was defined as the direct postoperative BMD (< 3 weeks). Data on BMD were collected for all Gruen zones (as mean BMD g/cm2 or percentual change in BMD, with or without standard deviation). The studies that presented their data in mg/cm2 or g/cm2 were converted into percentual change of baseline BMD in each Gruen zone.

Gruen zones are defined areas around the hip stem to assess osteolysis. Originally, they were divided into seven zones for conventional stems, but are also appliable for short stems (Fig. 1). Gruen 1 begins on the proximal lateral side (greater trochanter) and Gruen 7 ends on the proximal medial side (calcar region). If studies used a different Gruen zone numbering, this was converted to the zones as depicted in Fig. 1.

Fig. 1
Fig. 1

Gruen zones in short hip stems.

Citation: EFORT Open Reviews 6, 11; 10.1302/2058-5241.6.210030

Search strategy

To identify all suitable studies, the following search terms were used in Medline, PubMed and Embase: “hip arthroplasty OR hip replacement” and “short stem OR anatomic stem” and “dual-energy OR DEXA” between January 1990 and June 2020. More details of the search can be requested from the authors. The search was performed by two reviewers (PV and DH), independently. Language was restricted to English, German, French and Spanish. References of retrieved studies, potentially meeting the criteria, were also screened. Abstracts from congress, scientific meetings and unpublished reports were excluded.

Methods of review

Selection of studies

Studies were selected by reviewing the title and abstract to identify potentially relevant articles by two reviewers (PV and DH). Disagreement was resolved through discussion by the two reviewers, with a third reviewer (SW) as final vote with remaining discussions. The full text was retrieved and evaluated for definitive inclusion, when the title, keywords or abstract showed information on bone mineral density or DEXA in total hip arthroplasty.

Data extraction

Two authors (SW and JV) independently performed the data extraction. Predefined extraction forms were used to collect the following data from each study: author, year of publication, type of implant, manufacturer, whether the implant is still on the market, number of patients/hips, years of follow-up, and the patient demographics (such as age, gender and BMI). BMD data were collected in mg/cm2, g/cm2 or percentual change of BMD at baseline, all with the standard deviation if described. When available, additional Harris Hip Scores (HHS) were collected. In case of non-numerical presentation of BMD, the authors were contacted to retrieve numerical data.

All hip implants were defined as either short or anatomical hip stems. According to the classification system as used by van Oldenrijk et al21 and described by Falez et al,15 the short stems were classified as partial collum or trochanter sparing, and the anatomic stems as trochanter sparing or trochanter harming. Additionally, all stem types were categorized based on anchoring site as metaphyseal, metadiaphyseal or diaphyseal anchoring.22 Follow-up was categorized as 1–3 months, six months, one year, two years, 3–5 years and 5–10 years.

Methodological quality

Methodological quality was assessed by use of the Methodological Index for Non-Randomized Studies (MINORS) criteria for quality assessment of the studies23 by two authors (SW and PV) independently. Discrepancies were resolved by consensus.

Analysis

Descriptive statistics were used to present study details. BMD data were converted into overall weighted mean of bone loss, expressed in percentages, for short and anatomic stems. The standard error (SE) was calculated using the pooled standard deviations in percentages. A threshold of ten percent was defined by the authors as the clinically relevant cut-off value for bone loss in the proximal Gruen zones (G1 and G7); however, there is no known threshold. The review of Knutsen et al showed in different types of conventional hip stems a wide range of percentual bone loss, where after two years in non-cemented femoral stems there was bone loss of at least 10%. A hip stem was classified as bone stock preserving in case the bone loss was less than this threshold. Additionally, all short stems that are still on the market were selected and their course of bone loss during a two-year follow-up period was investigated.

Results

Selection procedure

In total, 681 references were found in the online databases, all published between January 1997 and June 2020. After removal, a total of 470 articles remained for reviewing titles, abstracts and keywords. Finally, 29 studies were included in this review (Fig. 2).

Fig. 2
Fig. 2

Flowchart with selection of included studies.

Note. BMD, bone mineral density; DEXA, dual-energy X-ray absorptiometry; CT, computerized tomography; FU, follow-up.

Citation: EFORT Open Reviews 6, 11; 10.1302/2058-5241.6.210030

Study characteristics

The 29 included studies contained a total of 2095 patients (2265 hips). Details of the included studies are presented in Table 1. The population consisted of 56% male and 44% female patients, with an average age of 55 years (range, 17–83 years) and a mean body mass index (BMI) of 28.4 (range, 16.9–49.6). Eighteen different stem designs were found, of which nine were short stems and nine anatomic stems (Table 2). According to the level of osteotomy, the stems were classified as trochanter harming (seven stems),2430 trochanter sparing (seven stems)25,26,3140 and partial collum (four stems).4150According to site of anchoring, the stems were classified as metaphyseal (three stems),25,26,36,41,43,44,46,47 metadiaphyseal (seven stems)24,26,3742,45,48,51 or diaphyseal (eight stems)25,2730,3336,52 anchoring. After patients who were lost to follow-up were excluded, 1796 hips remained to analyse. All included hip stems with their respective bone loss are depicted in Table 2. The bone loss in short and anatomic stems at one year follow-up is comparable in Gruen 7; however, there appears to be a difference in Gruen 1 at two-year follow-up (Fig. 3). The bone loss in all Gruen zones stabilized after three months, except for Gruen 7 which displayed a continuous bone loss until six months in the anatomic stems (Table 3).

Table 1.

Summary of included studies

Author, year Patients/hips (n) Age (mean) M/F

(n)
Body mass index (mean) Follow-up (years) Design MINORS
Arabmotlag, 200352 15/15 48 7/8 2 CTX 17/22
Boller, 201950 39/39 51 13/26 28 2 Metha 12/16
28/28 66 16/12 27 2 Metha
Brinkmann, 201541 24/24 59 12/12 27 1 Metha 16/22
26/26 60 16/10 27 Nanos
Ercan, 201642 62/62 57 28/34 29 1 Minihip 10/16
Fokter, 201524 19/19 60 8/11 28 1 Unibionix 16/22
Freitag, 201633 57/57 57 36/21 30 1 Fitmore 19/22
Gasbarra, 201434 33/33 62 15/18 24 1 Fitmore 16/22
Hayashi, 201670 21/21 68 11/56 23 2 Trilock 12/16
Jahnke, 201443 40/40 55 20/20 27 1 Metha 13/16
Kim, 201125 50/60 54 22/28 26 3 Proxima 16/22
50/60 52 24/26 25 Profile
Kim, 201626 201/221 53 118/83 30 10 Proxima 19/22
400/530 53 264/136 29 IPS
Kim, 201636 200/200 53 138/62 30 12 Proxima 19/22
200/200 53 138/62 30 Profile
Leali, 200427 10/10 65 4/6 3 Revelation 10/16
Lerch, 201244 25/25 59 16/9 25 2 Metha 13/16
Meyer, 201932* 54/54 5 Fitmore
Nysted, 201151 43/43 55 18/28 5 Unique 16/22
35/35 53 13/28 ABG-I
Panisello, 200937 56/56 60 27/29 28 5 ABG-I 19/22
54/54 59 26/28 27 ABG-II
Parchi, 201749 20/20 54 11/9 4 Metha 12/16
Rahmy, 200438 24/24 62 11/18 28 3 ABG-I 18/22
Salemyr, 201531 26/26 62 11/15 27 4 Proxima 21/22
Shafy, 201645 26/26 43 20/6 27 2 Minihip 12/16
Sluimer, 200628 40/40 53 15/25 2 Omnifit-HA 1090 18/22
Steens, 201546 20/20 49 12/8 26 5 ESKA cut 2000 14/16
Synder, 201547 36/36 50 18/18 1 Metha 12/16
Vd Wal, 200640 25/25 62 17/8 2 ABG-I 16/22
26/26 60 12/14 2 ABG-II
Vd Wal, 200839 24/24 60 11/13 2 ABG-I 15/22
White, 200829 27/27 37 14/13 25 5 Epoch 9/16
Wixson, 199730 35/35 50 16/19 2 Custom made 9/16
Zeh, 201348 25/25 60 15/10 29 1 Nanos 11/16

Note. MINORS, Methodological Index for Non-Randomized Studies.

*Study between two age groups.

**follow-up of Freitag et al, 201633.

Table 2.

Classification and course of bone loss per hip stem design from baseline in Gruen 1 and 7 (in %)

Gruen 1 Type Osteotomy Anchoring Studies (n) N 3M 6M 1Y 2Y
ABG-I3538,51 Anatomic Trochanter sparing Metadiaphyseal 5 164 –11 –11 –13 –12
ABG-II35,38 Anatomic Trochanter sparing Metadiaphyseal 2 80 –5 –7 –8 –9
Custom made Biomet30 Anatomic Trochanter harming Diaphyseal 1 31 –10 –17 –8
CTX52 Anatomic Trochanter sparing Diaphyseal 1 15 –2
Epoch29 Anatomic Trochanter harming Diaphyseal 1 12 –17 –19
ESKA cut 200046 Short Partial collum Metaphyseal 1 30 –3 –1
Fitmore31,32,40 Short Trochanter sparing Diaphyseal 3 90* –3 0 –4
IPS33 Anatomic Trochanter harming Metadiaphyseal 1 385 0
Metha41,43,44,47,49, 50 Short Partial collum Metaphyseal 6 206 –9 –9 –8 –5
Minihip42,45 Short Partial collum Metadiaphyseal 2 88 –11 –12 –8 –3
Nanos41,48 Short Partial collum Metadiaphyseal 2 51 –9 –14
Omnifit-HA 109028 Anatomic Trochanter harming Diaphyseal 1 35 –11 –13 –14 –13
Profile25,26 Anatomic Trochanter harming Diaphyseal 2 60 –27
Proxima25,26,34, 39 Short Trochanter sparing Metaphyseal 4 333 –6 –3 –1 –3
Revelation27 Anatomic Trochanter harming Diaphyseal 1 10 –1 –2 –4
Trilock33 Short Trochanter sparing Diaphyseal 1 65 –6 –5 –3
Unibionix24 Short Trochanter harming Metadiaphyseal 1 19 0 5 10
Unique51 Short Trochanter sparing Metadiaphyseal 1 43 –10 –9 –10 –11
Gruen 7 Type Osteotomy Anchoring Studies (n) N 3M 6M 1Y 2Y
ABG-I 3538,51 Anatomic Trochanter sparing Metadiaphyseal 5 164 –12 –22 –23 –15
ABG-II35,38 Anatomic Trochanter sparing Metadiaphyseal 2 80 –5 –10 –16 –12
Custom made Biomet30 Anatomic Trochanter harming Diaphyseal 1 31 –17 –19 15
CTX52 Anatomic Trochanter sparing Diaphyseal 1 15 –34
Epoch29 Anatomic Trochanter harming Diaphyseal 1 12 –10 –12
ESKA cut 200046 Short Partial collum Metaphyseal 1 30 –3 1
Fitmore31,32,40 Short Trochanter sparing Diaphyseal 3 90* –14 –6 –8
IPS33 Anatomic Trochanter harming Metadiaphyseal 1 385 –6
Metha41,43,44,47,49, 50 Short Partial collum Metaphyseal 6 206 –15 –11 –13 –3
Minihip42,45 Short Partial collum Metadiaphyseal 2 88 –11 –12 –9 –3
Nanos41,48 Short Partial collum Metadiaphyseal 2 51 –9 –8
Omnifit-HA 109028 Anatomic Trochanter harming Diaphyseal 1 35 –11 –13 –14 –13
Profile25,26 Anatomic Trochanter harming Diaphyseal 2 60 –33
Proxima25,26,34,39 Short Trochanter sparing Metaphyseal 4 333 –11 –12 –7 –13
Revelation27 Anatomic Trochanter harming Diaphyseal 1 10 –4 –5 –5
Trilock33 Short Trochanter sparing Diaphyseal 1 65 –9 –9 –14
Unibionix24 Short Trochanter harming Metadiaphyseal 1 19 –12 –12 –15
Unique51 Short Trochanter sparing Metadiaphyseal 1 43 –17 –20 –21 –24

*Number of patients of two studies (Freitag33 and Gasbarra34 as Meyer32 et al is a follow-up study of Freitag.33

Table 3.

Mean percentual change in bone mineral density in short and anatomic stems as a proportion from baseline (< 3 weeks postoperative)

Short Time (months) 3 6 12 24 36–60 60–120
Nr of hips (G1, G7) 397 (12) 456 (12) 855 (18) 266 (7) 335 (4) 221 (3)
Nr of hips (G2–G6) 371 (11) 430 (11) 582 (16) 240 (6) 117 (3) 63 (2)
Gruen 1 –7 –8 –5 –5 1 –2
Gruen 2 –7 –5 –4 –1 –9 –7
Gruen 3 –1 –1 0 1 –2 –4
Gruen 4 –4 –4 –2 –1 –5 –6
Gruen 5 –3 –2 –2 0 0 –3
Gruen 6 –5 –1 1 2 –4 0
Gruen 7 –12 –11 –10 –11 –13 –15
Anatomic Time (months) 3 6 12 24 36–60 60–120
Nr of hips (G1, G7) 155 (5) 245 (6) 731 (9) 227 (8) 504 (4) 631 (5)
Nr of hips (G2–G6) 155 (5) 245 (6) 257 (7) 227 (8) 59 (2) 157 (3)
Gruen 1 –9 –10 –8 –11 –13 –2
Gruen 2 –7 –4 –5 –6 –7 4
Gruen 3 –5 –4 –2 –2 –5 –1
Gruen 4 –4 –5 –1 –1 –2 –1
Gruen 5 –4 –1 –2 –1 –1 3
Gruen 6 –6 –4 –4 –3 –5 3
Gruen 7 –10 –17 –14 –15 –13 –25

Between brackets number of studies.

Fig. 3
Fig. 3

Percentual bone loss per type of hip stem in Gruen 1 and 7, with pooled standard error.

Note. BMD, bone mineral density.

Citation: EFORT Open Reviews 6, 11; 10.1302/2058-5241.6.210030

Short stem

A total number of 1068 patients (1098 hips) were included in 19 studies (not including Meyer et al,32 of whom 53.7% were male (n = 574/1068) and 46.3% were female (n = 494/1068). The average age was 55 years with a mean BMI of 25. Once patients who were lost to follow-up were excluded, a total of 915 hips remained. A mean total value in HHS of 94 at a minimal follow-up of one year (range, 1–10 years) was reported.

Gruen 1 showed a bone loss of 5% after one year (n = 855) and 5% after two years (n = 266). Gruen 7 was the most affected zone, with a bone loss of 10% after one year and 11% after two years (Table 3).

Anatomic stem

A total of 1027 patients (1167 hips) were included in 13 studies, of whom 56% were male (n = 597/1027) and 44% female (n = 430/1027). In this population, the mean age was 54 years with a mean BMI of 23. Lost to follow-up patients were excluded, leaving a total of 881 hips in this subgroup. The HHS had a mean total value of 96 points at a minimal follow-up of one year (range, 1–10 years).

Gruen 1 showed bone loss of 8% after one year (n = 731) and 11% after two years (n = 227). Bone loss in Gruen 7 was 14% after one year and 15% after two years (Table 3).

Osteotomy levels (Fig. 4)

Bone loss in Gruen 1 was, in both trochanter sparing and partial collum hip stems, less than 10%, with a slightly better preservation in trochanter sparing hip stems at one year follow-up. This was not seen at two-year follow-up. A difference of > 10% bone loss was found at two-year follow-up between trochanter sparing and partial collum hip stems in Gruen 7.

Fig. 4
Fig. 4

Percentual bone loss in type of osteotomy in Gruen 1 and 7 (short and anatomic), with pooled standard error.

Note. BMD, bone mineral density.

Citation: EFORT Open Reviews 6, 11; 10.1302/2058-5241.6.210030

In anatomic stems, the trochanter harming stems were bone stock preserving at one-year follow-up in Gruen 1, but this preservation was not seen at two-year follow-up. There is a difference between trochanter sparing stems in short and anatomic stems, in favour of the short stems.

Anchoring types (Fig. 5)

In short stems, metaphyseal anchoring was bone stock preserving in the two Gruen zones at one- and two-year follow-up. Metadiaphyseal anchoring also achieves this in Gruen 1, but not in Gruen 7. Comparing metadiaphyseal short stems and anatomic stems, both stems had equal bone loss in Gruen 1 and 7.

Fig. 5
Fig. 5

Percentual bone loss in type of anchoring in Gruen 1 and 7 (short and anatomic), with pooled standard error.

Note. BMD, bone mineral density.

Citation: EFORT Open Reviews 6, 11; 10.1302/2058-5241.6.210030

Short stem types (Fig. 6)

The progress of the different types of short stems that are still on the market in Gruen 1 and Gruen 7 over time varies. The total mean percentual changes in BMD of the short and anatomic stems were included as reference line, with the threshold of 10%.

Fig. 6
Fig. 6

Bone loss in % per short stem design (still on the market) in Gruen 1 and 7 over time, with the total mean per stem included. Threshold of 10% bone loss added.

Note. BMD, bone mineral density.

Citation: EFORT Open Reviews 6, 11; 10.1302/2058-5241.6.210030

At two years follow-up, the Metha (Braun, Hessen, Germany), Fitmore (ZimmerBiomet, Warsaw, USA) and Minihip (Corin group PLC, Circencester, United Kingdom) were bone stock preserving in Gruen 1 and 7. The Trilock (DePuy, Warsaw, IN) was bone stock preserving in Gruen 1,but this was not achieved in Gruen 7. The Nanos (Smith & Nephew, Marl, Germany) is bone stock preserving in Gruen 7 at one-year follow-up, but did have more than 10% bone loss in Gruen 1.

Discussion

This review compiled the available data on periprosthetic bone loss in short and anatomic hip stems and defined which type of stem is best capable of preserving bone stock. Overall one-year follow-up results revealed that short and anatomic stems are bone stock preserving in Gruen 1. Furthermore, only the short stem was bone stock preserving in Gruen 7 at one-year follow-up. In the rest of the follow-up moments, in both short and anatomic stems, there was more than 10% bone loss. Bone stock was preserved in the remaining Gruen zones in short and anatomic stems.

This review was restricted to short stems and anatomic stems; hip resurfacing stems were not included due to different physical loading and anchoring principles. Uncemented conventional hip stems were covered in the review by Knutsen et al.13 All measurements were obtained with DEXA scans, a widely used and reliable technique for measurement of periprosthetic bone loss.17,53 The focus of interest in this review were Gruen 1and 7, as these two zones are most representative for proximal loading of the femur and where periprosthetic bone loss is mainly observed.13,54,55 These two zones are also the most comparable between different stem designs, as stem length defines the length of the Gruen zone for Gruen 2 until Gruen 6, but is less important for Gruen 7 (the calcar/lesser trochanter region) and Gruen 1 (greater trochanter region).

By maintaining a greater part of the postoperative physical loading in the metaphysis, the trochanter region has less bone stock loss in short (-5%) and anatomic (-8%) stems one-year postoperatively than uncemented conventional stems (–9.8%), as shown in the review by Knutsen et al.13 In the calcar region it is more difficult to preserve bone stock in uncemented conventional stems (–19%).13 Short stems (-10%) show better bone stock preservation one year after implantation compared to uncemented conventional stems. The more proximal anchoring anatomic stems (-14%) also provide a smaller amount of bone loss than uncemented conventional stems in the calcar region.

The present study had some limitations. This review has focused only on the classification of the hip stem, thereby neglecting implant-specific characteristics such as geometry, coating, stiffness and material that are also associated with BMD.13 The IPS (anatomic) and Unibionix stem (short) both show remarkable bone stock preservation in Gruen 1, the Unique stem (short) displays bone loss of more than 20% in Gruen 7 and the Profile stem (anatomic) has bone loss of more than 20% in both Gruen zones. These results are not in line with other results of the same type of hip stems in their category, indicating that other implant-specific characteristics play an important role. Unfortunately, stem positioning and osteotomy level were not mentioned in most studies, therefore their effect on BMD could not be assessed in this review. But it could be an explanation between two different results in the same type of hip stem (Metha), where one study describes bone loss of 23% in Gruen 7 and another 8%. A study by Brinkmann et al56 showed a different straining distribution in stems which were placed in a varus or valgus alignment.

Most studies reported data on BMD direct postoperatively in both Gruen zones 1 and 7; however, 19 studies did not. They compared follow-up BMD to the preoperative BMD (n = 5), the BMD of the contralateral side (n = 3), the BMD value without Gruen zones (n = 4) or made no comparison at all (n = 7). Therefore, these studies were excluded from this review. A threshold of 10% bone loss was chosen as clinically relevant in this review as reference value for all short hip stems, thereby making a stem bone stock preserving when it shows less than 10% bone loss. However, there is no known threshold for how much bone loss is of clinical significance for implant failure. Long-term survival in hip stems can be affected by bone loss in Gruen 1 and 7, but in case of sufficient anchoring in the other (more distal) Gruen zones, implant stability can be maintained. As conventional stems have a higher percentage of bone loss, most stems also have excellent long-term survival, as seen with the 13A Orthopaedic Data Evaluation Panel (ODEP) rating for the CLS and Exeter for example. However, if the femoral component fails in the long term due to aseptic loosening, this is often accompanied with (proximal) bone stock loss and mostly seen in Gruen 1 and Gruen 7. As revision surgery in total hip arthroplasty is rising, due to a higher incidence of younger and more active patients with total hip arthroplasty,5761 this loss can pose more difficulty during revision surgery62 and influence the type of hip stem as femoral replacement.63 The chance of complications, such as fractures, and the difficulty of the revision is theoretical lower when there is still sufficient bone stock.64 Especially in patients who have a chance of re-revision in THA, maintaining bone stock seems to play an important role.65

The majority of the studies concerned follow-up data limited to the first postoperative year. Since bone activity is a continuous process, medium- and long-term bone remodelling is important for better understanding of the effect that short stems have on the physical loading.26,66,67 Especially since the patient population is in a stage of life where bone stock will naturally decline due to age, reduced activity and hormonal shifts.68 Nevertheless, the promising course of bone remodelling in this review around the short stem during the first two postoperative years could yield information on bone stock preservation in the long term.12,69

Both classification systems based on osteotomy level as well as anchoring site revealed varying amounts of bone loss within their classification among short and anatomical stems. This could indicate that positioning of the stem can play a role in bone remodelling.56,70 For instance, the Nanos stem can be placed in a varus or valgus position, depending on the hip anatomy of the individual patient in terms of the femoral offset. Osteotomy level can therefore be either partial collum or trochanter sparing while maintaining the anchoring site. Classification on just osteotomy level could not be accurate enough. Therefore, classification according to type of anchoring should be described in further research when describing the effect of the hip stem on femoral bone loss.

Conclusion

This study contributes to the understanding of periprosthetic bone mineral density in short stem hip arthroplasty. Short stems are capable of preserving bone stock by maintaining the postoperative physical loading in the metaphysis. Though there is still loss of bone, this is slightly reduced in short stems compared to anatomic stems. Whether this effect is clinically relevant and remains in the long term is yet to be established.

Open access

This article is distributed under the terms of the Creative Commons Attribution-Non Commercial 4.0 International (CC BY-NC 4.0) licence (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed.

ICMJE Conflict of interest statement

The author declares no conflict of interest relevant to this work.

Funding statement

No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.

Social media

[Twitter: SheryldeWaard, Linkedin: Sheryl de Waard]

OA licence text

This article is distributed under the terms of the Creative Commons Attribution-Non Commercial 4.0 International (CC BY-NC 4.0) licence (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed.

References

  • 1.

    Arthroplasty in the Picture Annual Report of the Dutch Arthroplasty Register (Landelijke Registratie Orthopedische Implantaten) 2014. www.lroi-rapportage.nl (date last accessed 21 October 2021).

    • PubMed
    • Export Citation
  • 2.

    Landelijke Registratie Orthopedische Implantaten (LROI) ©2016 Online LROI-Rapportage 2015. www.lroi-rapportage.nl (date last accessed 21 October 2021).

    • PubMed
    • Export Citation
  • 3.

    Swarup I, Marshall AC, Lee YY, Figgie MP. Implant survival and patient-reported outcomes after total hip arthroplasty in young patients with developmental dysplasia of the hip. Hip Int 2017;26:367373.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 4.

    Swarup I, Sutherland R, Jayme C, Burket JC, Figgie MP. Total hip arthroplasty in young patients with post-traumatic arthritis of the hip. Hip Int 2017;27:546550.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 5.

    Swarup I, Shields M, Mayer EN, Hendow CJ, Burket JC, Figgie MP. Outcomes after total hip arthroplasty in young patients with osteonecrosis of the hip. Hip Int 2017;27:286292.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 6.

    Santaguida PL, Hawker GA, Hudak PL, Glazier R & Mahomed NNet al. Patient characteristics affecting the prognosis of total hip and knee joint arthroplasty: a systematic review. Can J Surg 2008;51:428436.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 7.

    Alm JJ, Mäkinen TJ, Lankinen P, Moritz N, Vahlberg T, Aro HT. Female patients with low systemic BMD are prone to bone loss in Gruen zone 7 after cementless total hip arthroplasty. Acta Orthop 2009;80:531537.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 8.

    Sköldenberg OG, Bodén HSG, Salemyr MO, Ahl TE, Adolphson PY. Periprosthetic proximal bone loss after uncemented hip arthroplasty is related to stem size: DXA measurements in 138 patients followed for 2–7 years. Acta Orthop 2006;77:386392.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 9.

    Rosenthall L, Bobyn JD, Tanzer M. Bone densitometry: influence of prosthetic design and hydroxyapatite coating on regional adaptive bone remodelling. Int Orthop 1999;23:325329.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 10.

    Tanzer M, Kantor S, Rosenthall L, Bobyn JD. Femoral remodeling after porous-coated total hip arthroplasty with and without hydroxyapatite-tricalcium phosphate coating: a prospective randomized trial. J Arthroplasty 2001;16:552558.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 11.

    Parchi PD, Cervi V & Piolanti Net al. Densitometric evaluation of periprosthetic bone remodeling. Clin Cases Miner Bone Metab 2014;11:226231.

  • 12.

    Roth A, Richartz G & Sander Ket al. Verlauf der periprothetischen Knochendichte nach Hüfttotalendoprothesenimplantation. Abhängigkeit von Prothesentyp und knöcherner Ausgangssituation. Orthopade 2005;34:334344.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 13.

    Knutsen AR, Lau N, Longjohn DB, Ebramzadeh E, Sangiorgio SN. Periprosthetic femoral bone loss in total hip arthroplasty: systematic analysis of the effect of stem design. Hip Int 2017;27:2634.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 14.

    Albanese CV, Santori FS, Pavan L, Learmonth ID, Passariello R. Periprosthetic DXA after total hip arthroplasty with short vs. ultra-short custom-made femoral stems: 37 patients followed for 3 years. Acta Orthop 2009;80:291297.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 15.

    Falez F, Casella F, Papalia M. Current concepts, classification, and results in short stem hip arthroplasty. Orthopedics 2015;38:S6S13.

  • 16.

    Cohen B, Rushton N. Accuracy of DEXA measurement of bone mineral density after total hip arthroplasty. J Bone Joint Surg [Br] 1995;77-B:479483.

  • 17.

    Bloebaum RD, Liau DW, Lester DK, Rosenbaum TG. Dual-energy X-ray absorptiometry measurement and accuracy of bone mineral after unilateral total hip arthroplasty. J Arthroplasty 2006;21:612622.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 18.

    Martini F, Lebherz C, Mayer F, Leichtle U, Kremling E, Sell S. Precision of the measurements of periprosthetic bone mineral density in hips with a custom-made femoral stem. J Bone Joint Surg [Br] 2000;82-B:10651071.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 19.

    Kroger H, Miettinen H, Arnala I, Koski E, Rushton N, Suomalainen O.Evaluation of periprosthetic bone using dual-energy X-ray absorptiometry: precision of the method and effect of operation on bone mineral density. J Bone Miner Res 1996;11:15261530.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 20.

    Gruen T, McNeice G, Amstutz H. ‘Modes of failure’ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 1979;141:1727.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 21.

    van Oldenrijk J, Molleman J, Klaver M, Poolman RW, Haverkamp D. Revision rate after short-stem total hip arthroplasty: a systematic review of 49 studies. Acta Orthop 2014;85:250258.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 22.

    McTighe T, Keggi J, David Stulberg S, Keppler L, Brazil D, McPherson EJ. Total hip stem classification system. Reconstr Rev 2014;4:2428.

  • 23.

    Slim K, Nini E, Forestier D, Kwiatkowski F, Panis Y, Chipponi J. Methodological Index for Non-Randomized Studies (MINORS): development and validation of a new instrument. ANZ J Surg 2003;73:712716.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 24.

    Fokter SK, Sarler T, Strahovnik A, Repše-Fokter A. Results of total hip arthroplasty using a bionic hip stem. Int Orthop 2015;39:10651071.

  • 25.

    Kim YH, Choi Y, Kim JS. Comparison of bone mineral density changes around short, metaphyseal-fitting, and conventional cementless anatomical femoral components. J Arthroplasty 2011;26:931940.e1.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 26.

    Kim YH, Park JW, Kim JS. Metaphyseal engaging short and ultra-short anatomic cementless stems in young and active patients. J Arthroplasty 2016;31:180185.

  • 27.

    Leali A, Fetto JF. Preservation of femoral bone mass after total hip replacements with a lateral flare stem. Int Orthop 2004;28:151154.

  • 28.

    Sluimer JC, Hoefnagels NHM, Emans PJ, Kuijer R, Geesink RGT. Comparison of two hydroxyapatite-coated femoral stems: clinical, functional, and bone densitometry evaluation of patients randomized to a regular or modified hydroxyapatite-coated stem aimed at proximal fixation. J Arthroplasty 2006;21:344352.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 29.

    White SP, Lee MB, Galpin J, Learmonth ID. Ten-year results of a bone-preserving low-modulus composite total hip replacement stem. Hip Int 2008;18:8187.

  • 30.

    Wixson RL, Stulberg SD, Van Flandern GJ, Puri L. Maintenance of proximal bone mass with an uncemented femoral stem analysis with dual-energy x-ray absorptiometry. J Arthroplasty 1997;12:365372.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 31.

    Salemyr M, Muren O & Ahl Tet al. Lower periprosthetic bone loss and good fixation of an ultra-short stem compared to a conventional stem in uncemented total hip arthroplasty. Acta Orthop 2015;86:659666.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 32.

    Meyer JS, Freitag T, Reichel H, Bieger R. Periprosthetic bone mineral density changes after implantation of a curved bone preserving hip stem compared to a standard length straight stem: 5-yr results of a prospective, randomized DXA-analysis. J Clin Densitom 2019;22:96103.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 33.

    Freitag T, Hein MA, Wernerus D, Reichel H, Bieger R. Bone remodelling after femoral short stem implantation in total hip arthroplasty: 1-year results from a randomized DEXA study. Arch Orthop Trauma Surg 2016;136:125130.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 34.

    Gasbarra E, Celi M & Perrone FLet al. Osseointegration of Fitmore stem in total hip arthroplasty. J Clin Densitom 2014;17:307313.

  • 35.

    Hayashi S, Nishiyama T & Fujishiro Tet al. Maintenance of periprosthetic bone mineral density with a cementless three dimensional straight tapered stem. Hip Int 2011;21:163167.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 36.

    Kim YH, Park JW, Kim JS. Ultrashort versus conventional anatomic cementless femoral stems in the same patients younger than 55 years. Clin Orthop Relat Res 2016;474:20082017.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 37.

    Panisello JJ, Canales V, Herrero L, Herrera A, Mateo J, Caballero MJ. Changes in periprosthetic bone remodelling after redesigning an anatomic cementless stem. Int Orthop 2009;33:373379.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 38.

    Rahmy AIA, Gosens T, Blake GM, Tonino A, Fogelman I. Periprosthetic bone remodelling of two types of uncemented femoral implant with proximal hydroxyapatite coating: a 3-year follow-up study addressing the influence of prosthesis design and preoperative bone density on periprosthetic bone loss. Osteoporos Int 2004;15:281289.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 39.

    van der Wal BCH, Rahmy AIA, Grimm B, Heyligers IC, Tonino AJ. Preoperative bone quality as a factor in dual-energy X-ray absorptiometry analysis comparing bone remodelling between two implant types. Int Orthop 2008;32:3945.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 40.

    van der Wal BCH, Rahmy AIA, Grimm B, Blake GM, Heyligers IC, Tonino AJ. The influence of implant design on periprosthetic bone remodelling of two types hip stems. Hip Int 2006;16:817.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 41.

    Brinkmann V, Radetzki F, Delank KS, Wohlrab D, Zeh A. A prospective randomized radiographic and dual-energy X-ray absorptiometric study of migration and bone remodeling after implantation of two modern short-stemmed femoral prostheses. J Orthop Traumatol 2015;16:237243.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 42.

    Ercan A, Sokkar SM, Schmid G, Filler TJ, Abdelkafy A, Jerosch J. Periprosthetic bone density changes after MiniHip TM cementless femoral short stem: one-year results of dual-energy X-ray absorptiometry study. SICOT J 2016;2:40.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 43.

    Jahnke A, Engl S & Altmeyer Cet al. Changes of periprosthetic bone density after a cementless short hip stem: a clinical and radiological analysis. Int Orthop 2014;38:20452050.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 44.

    Lerch M, von der Haar-Tran A, Windhagen H, Behrens BA, Wefstaedt P, Stukenborg-Colsman CM. Bone remodelling around the Metha short stem in total hip arthroplasty: a prospective dual-energy X-ray absorptiometry study. Int Orthop 2012;36:533538.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 45.

    Shafy TA, Sayed A, Abdelazeem AH. Study of the bone behavior around a neck preserving short stem implant: bone densitometric analysis over a span of two years. SICOT J 2016;2:31.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 46.

    Steens W, Boettner F, Bader R, Skripitz R, Schneeberger A. Bone mineral density after implantation of a femoral neck hip prosthesis: a prospective 5 year follow-up. BMC Musculoskelet Disord 2015;16:192.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 47.

    Synder M, Krajewski K, Sibinski M, Drobniewski M. Periprosthetic bone remodeling around short stem. Orthopedics 2015;38:S40S45.

  • 48.

    Zeh A, Pankow F, Röllinhoff M, Delank S, Wohlrab D. A prospective dual-energy X-ray absorptiometry study of bone remodeling after implantation of the Nanos short-stemmed prosthesis. Acta Orthop Belg 2013;79:174180.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 49.

    Parchi PD, Ciapini G, Castellini I, Mannucci C, Nucci A-M, Piolanti N, Maffei S, Lisanti M. Evaluation of the effects of the Metha® short stem on periprosthetic bone remodelling in total hip arthroplasties: results at 48 months. Surg Technol Int 2017;30:346351.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 50.

    Boller S, Jahnke A, Augustin L, Ahmed G, Rickert M, Alexander B. Age-related osseointegration of a short hip stem: a clinical and radiological 24 months follow-up. Arch Orthop Trauma Surg 2019;139:405410.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 51.

    Nysted M, Benum P, Klaksvik J, Foss O, Aamodt A. Periprosthetic bone loss after insertion of an uncemented, customized femoral stem and an uncemented anatomical stem: a randomized DXA study with 5-year follow-up. Acta Orthop 2011;82:410416.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 52.

    Arabmotlagh M, Hennigs T, Rittmeister M. Femoral periprosthetic bone remodelling after implantation of custom made anatomic and standard straight stem hip prostheses. Z Orthop Ihre Grenzgeb 2003;141:519525.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 53.

    Cohen B, Rushton N. Accuracy of DEXA measurement of bone mineral density after total hip arthroplasty. J Bone Joint Surg [Br] 1995;77-B:77479.

  • 54.

    Grochola LF, Habermann B, Mastrodomenico N, Kurth A. Comparison of periprosthetic bone remodelling after implantation of anatomic and straight stem prostheses in total hip arthroplasty. Arch Orthop Trauma Surg 2008;128:383392.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 55.

    Boyle C, Kim IY. Comparison of different hip prosthesis shapes considering micro-level bone remodeling and stress-shielding criteria using three-dimensional design space topology optimization. J Biomech 2011;44:17221728.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 56.

    Brinkmann V, Radetzki F, Gutteck N, Delank S, Zeh A. Influence of varus/valgus positioning of the Nanos® and Metha® short-stemmed prostheses on stress shielding of metaphyseal bone. Acta Orthop Belg 2017;83:5766.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 57.

    Duncan CP, Masterson EL, Masri BA. Impaction allografting with cement for the management of femoral bone loss. Orthop Clin North Am 1998;29:297305.

  • 58.

    Goff TAJ, Bobak P. Femoral impaction allografting for significant bone loss in revision hip arthroplasty. Hip Int 2017;27:281285.

  • 59.

    Abdel MP, Roth PV, Harmsen WS, Berry DJ. What is the lifetime risk of revision for patients undergoing total hip arthroplasty? A 40-year observational study of patients treated with the Charnley cemented total hip arthroplasty. J Bone Joint Surg [Br] 2016;98-B:14361440.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 60.

    Cnudde PHJ, Nemes S, Bülow E, Timperley AJ, Whitehouse SL & Kärrholm Jet al. Risk of further surgery on the same or opposite side and mortality after primary total hip arthroplasty: a multi-state analysis of 133,654 patients from the Swedish Hip Arthroplasty Register. Acta Orthop 2018;89:386393.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 61.

    Kärrholm J, Mohaddes M, Odin D, Vinblad J, Rogmark C, Rolfson O. Swedish Hip Arthroplasty Register annual report 2017. Gothenburg: Swedish Hip Arthroplasty Register, 2018.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 62.

    DeRogatis MJ, Issack PS. Total femoral replacement as a salvage operation for the treatment of massive femoral bone loss during revision total hip arthroplasty. JBJS Rev 2018;6:e9.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 63.

    Sculco PK, Abdel MP, Lewallen DG. Management of femoral bone loss in revision total hip arthroplasty. Hip Int 2015;25:380387.

  • 64.

    Glassman AH, Bobyn JD, Tanzer M. New femoral designs: do they influence stress shielding? Clin Orthop Relat Res 2006;453:6474.

  • 65.

    Sheth NP, Melnic CM, Rozell JC, Paprosky WG. Management of severe femoral bone loss in revision total hip arthroplasty. Orthop Clin North Am 2015;46:329342, ix.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 66.

    Götze C, Ehrenbrink J, Ehrenbrink H. Bleibt der Krafteinfluss der Kurzschaftprothese auf den methaphysären proximalen Femur begrenzt? Osteodensitometrische Analysen der NANOS®-Schaftendoprothese. Z Orthop Unfall 2010;148:398405.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 67.

    Thorey F, Hoefer C, Abdi-Tabari N, Lerch M, Budde S, Windhagen H. Clinical results of the Metha short hip stem: a perspective for younger patients? Orthop Rev (Pavia) 2013;5:e34.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 68.

    Boskey AL, Coleman R. Critical reviews in oral biology and medicine. J Dent Res 2010;89:13331348.

  • 69.

    Venesmaa PK, Ger HPJK, Miettinen HJA, Jurvelin JS, Suomalainen OT, Alhava EM. Monitoring of periprosthetic BMD after uncemented total hip arthroplasty with dual-energy X-ray absorptiometry: a 3-year follow-up study. J Bone Miner Res 2001;16:10561061.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 70.

    Hayashi S, Hashimoto S, Kanzaki N, Kuroda R, Kurosaka M. Stem anteversion affects periprosthetic bone mineral density after total hip arthroplasty. Hip Int 2016;26:260264.

    • PubMed
    • Search Google Scholar
    • Export Citation

 

  • Collapse
  • Expand
  • Fig. 1

    Gruen zones in short hip stems.

  • Fig. 2

    Flowchart with selection of included studies.

    Note. BMD, bone mineral density; DEXA, dual-energy X-ray absorptiometry; CT, computerized tomography; FU, follow-up.

  • Fig. 3

    Percentual bone loss per type of hip stem in Gruen 1 and 7, with pooled standard error.

    Note. BMD, bone mineral density.

  • Fig. 4

    Percentual bone loss in type of osteotomy in Gruen 1 and 7 (short and anatomic), with pooled standard error.

    Note. BMD, bone mineral density.

  • Fig. 5

    Percentual bone loss in type of anchoring in Gruen 1 and 7 (short and anatomic), with pooled standard error.

    Note. BMD, bone mineral density.

  • Fig. 6

    Bone loss in % per short stem design (still on the market) in Gruen 1 and 7 over time, with the total mean per stem included. Threshold of 10% bone loss added.

    Note. BMD, bone mineral density.

  • 1.

    Arthroplasty in the Picture Annual Report of the Dutch Arthroplasty Register (Landelijke Registratie Orthopedische Implantaten) 2014. www.lroi-rapportage.nl (date last accessed 21 October 2021).

    • PubMed
    • Export Citation
  • 2.

    Landelijke Registratie Orthopedische Implantaten (LROI) ©2016 Online LROI-Rapportage 2015. www.lroi-rapportage.nl (date last accessed 21 October 2021).

    • PubMed
    • Export Citation
  • 3.

    Swarup I, Marshall AC, Lee YY, Figgie MP. Implant survival and patient-reported outcomes after total hip arthroplasty in young patients with developmental dysplasia of the hip. Hip Int 2017;26:367373.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 4.

    Swarup I, Sutherland R, Jayme C, Burket JC, Figgie MP. Total hip arthroplasty in young patients with post-traumatic arthritis of the hip. Hip Int 2017;27:546550.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 5.

    Swarup I, Shields M, Mayer EN, Hendow CJ, Burket JC, Figgie MP. Outcomes after total hip arthroplasty in young patients with osteonecrosis of the hip. Hip Int 2017;27:286292.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 6.

    Santaguida PL, Hawker GA, Hudak PL, Glazier R & Mahomed NNet al. Patient characteristics affecting the prognosis of total hip and knee joint arthroplasty: a systematic review. Can J Surg 2008;51:428436.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 7.

    Alm JJ, Mäkinen TJ, Lankinen P, Moritz N, Vahlberg T, Aro HT. Female patients with low systemic BMD are prone to bone loss in Gruen zone 7 after cementless total hip arthroplasty. Acta Orthop 2009;80:531537.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 8.

    Sköldenberg OG, Bodén HSG, Salemyr MO, Ahl TE, Adolphson PY. Periprosthetic proximal bone loss after uncemented hip arthroplasty is related to stem size: DXA measurements in 138 patients followed for 2–7 years. Acta Orthop 2006;77:386392.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 9.

    Rosenthall L, Bobyn JD, Tanzer M. Bone densitometry: influence of prosthetic design and hydroxyapatite coating on regional adaptive bone remodelling. Int Orthop 1999;23:325329.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 10.

    Tanzer M, Kantor S, Rosenthall L, Bobyn JD. Femoral remodeling after porous-coated total hip arthroplasty with and without hydroxyapatite-tricalcium phosphate coating: a prospective randomized trial. J Arthroplasty 2001;16:552558.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 11.

    Parchi PD, Cervi V & Piolanti Net al. Densitometric evaluation of periprosthetic bone remodeling. Clin Cases Miner Bone Metab 2014;11:226231.

  • 12.

    Roth A, Richartz G & Sander Ket al. Verlauf der periprothetischen Knochendichte nach Hüfttotalendoprothesenimplantation. Abhängigkeit von Prothesentyp und knöcherner Ausgangssituation. Orthopade 2005;34:334344.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 13.

    Knutsen AR, Lau N, Longjohn DB, Ebramzadeh E, Sangiorgio SN. Periprosthetic femoral bone loss in total hip arthroplasty: systematic analysis of the effect of stem design. Hip Int 2017;27:2634.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 14.

    Albanese CV, Santori FS, Pavan L, Learmonth ID, Passariello R. Periprosthetic DXA after total hip arthroplasty with short vs. ultra-short custom-made femoral stems: 37 patients followed for 3 years. Acta Orthop 2009;80:291297.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 15.

    Falez F, Casella F, Papalia M. Current concepts, classification, and results in short stem hip arthroplasty. Orthopedics 2015;38:S6S13.

  • 16.

    Cohen B, Rushton N. Accuracy of DEXA measurement of bone mineral density after total hip arthroplasty. J Bone Joint Surg [Br] 1995;77-B:479483.

  • 17.

    Bloebaum RD, Liau DW, Lester DK, Rosenbaum TG. Dual-energy X-ray absorptiometry measurement and accuracy of bone mineral after unilateral total hip arthroplasty. J Arthroplasty 2006;21:612622.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 18.

    Martini F, Lebherz C, Mayer F, Leichtle U, Kremling E, Sell S. Precision of the measurements of periprosthetic bone mineral density in hips with a custom-made femoral stem. J Bone Joint Surg [Br] 2000;82-B:10651071.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 19.

    Kroger H, Miettinen H, Arnala I, Koski E, Rushton N, Suomalainen O.Evaluation of periprosthetic bone using dual-energy X-ray absorptiometry: precision of the method and effect of operation on bone mineral density. J Bone Miner Res 1996;11:15261530.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 20.

    Gruen T, McNeice G, Amstutz H. ‘Modes of failure’ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 1979;141:1727.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 21.

    van Oldenrijk J, Molleman J, Klaver M, Poolman RW, Haverkamp D. Revision rate after short-stem total hip arthroplasty: a systematic review of 49 studies. Acta Orthop 2014;85:250258.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 22.

    McTighe T, Keggi J, David Stulberg S, Keppler L, Brazil D, McPherson EJ. Total hip stem classification system. Reconstr Rev 2014;4:2428.

  • 23.

    Slim K, Nini E, Forestier D, Kwiatkowski F, Panis Y, Chipponi J. Methodological Index for Non-Randomized Studies (MINORS): development and validation of a new instrument. ANZ J Surg 2003;73:712716.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 24.

    Fokter SK, Sarler T, Strahovnik A, Repše-Fokter A. Results of total hip arthroplasty using a bionic hip stem. Int Orthop 2015;39:10651071.

  • 25.

    Kim YH, Choi Y, Kim JS. Comparison of bone mineral density changes around short, metaphyseal-fitting, and conventional cementless anatomical femoral components. J Arthroplasty 2011;26:931940.e1.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 26.

    Kim YH, Park JW, Kim JS. Metaphyseal engaging short and ultra-short anatomic cementless stems in young and active patients. J Arthroplasty 2016;31:180185.

  • 27.

    Leali A, Fetto JF. Preservation of femoral bone mass after total hip replacements with a lateral flare stem. Int Orthop 2004;28:151154.

  • 28.

    Sluimer JC, Hoefnagels NHM, Emans PJ, Kuijer R, Geesink RGT. Comparison of two hydroxyapatite-coated femoral stems: clinical, functional, and bone densitometry evaluation of patients randomized to a regular or modified hydroxyapatite-coated stem aimed at proximal fixation. J Arthroplasty 2006;21:344352.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 29.

    White SP, Lee MB, Galpin J, Learmonth ID. Ten-year results of a bone-preserving low-modulus composite total hip replacement stem. Hip Int 2008;18:8187.

  • 30.

    Wixson RL, Stulberg SD, Van Flandern GJ, Puri L. Maintenance of proximal bone mass with an uncemented femoral stem analysis with dual-energy x-ray absorptiometry. J Arthroplasty 1997;12:365372.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 31.

    Salemyr M, Muren O & Ahl Tet al. Lower periprosthetic bone loss and good fixation of an ultra-short stem compared to a conventional stem in uncemented total hip arthroplasty. Acta Orthop 2015;86:659666.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 32.

    Meyer JS, Freitag T, Reichel H, Bieger R. Periprosthetic bone mineral density changes after implantation of a curved bone preserving hip stem compared to a standard length straight stem: 5-yr results of a prospective, randomized DXA-analysis. J Clin Densitom 2019;22:96103.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 33.

    Freitag T, Hein MA, Wernerus D, Reichel H, Bieger R. Bone remodelling after femoral short stem implantation in total hip arthroplasty: 1-year results from a randomized DEXA study. Arch Orthop Trauma Surg 2016;136:125130.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 34.

    Gasbarra E, Celi M & Perrone FLet al. Osseointegration of Fitmore stem in total hip arthroplasty. J Clin Densitom 2014;17:307313.

  • 35.

    Hayashi S, Nishiyama T & Fujishiro Tet al. Maintenance of periprosthetic bone mineral density with a cementless three dimensional straight tapered stem. Hip Int 2011;21:163167.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 36.

    Kim YH, Park JW, Kim JS. Ultrashort versus conventional anatomic cementless femoral stems in the same patients younger than 55 years. Clin Orthop Relat Res 2016;474:20082017.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 37.

    Panisello JJ, Canales V, Herrero L, Herrera A, Mateo J, Caballero MJ. Changes in periprosthetic bone remodelling after redesigning an anatomic cementless stem. Int Orthop 2009;33:373379.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 38.

    Rahmy AIA, Gosens T, Blake GM, Tonino A, Fogelman I. Periprosthetic bone remodelling of two types of uncemented femoral implant with proximal hydroxyapatite coating: a 3-year follow-up study addressing the influence of prosthesis design and preoperative bone density on periprosthetic bone loss. Osteoporos Int 2004;15:281289.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 39.

    van der Wal BCH, Rahmy AIA, Grimm B, Heyligers IC, Tonino AJ. Preoperative bone quality as a factor in dual-energy X-ray absorptiometry analysis comparing bone remodelling between two implant types. Int Orthop 2008;32:3945.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 40.

    van der Wal BCH, Rahmy AIA, Grimm B, Blake GM, Heyligers IC, Tonino AJ. The influence of implant design on periprosthetic bone remodelling of two types hip stems. Hip Int 2006;16:817.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 41.

    Brinkmann V, Radetzki F, Delank KS, Wohlrab D, Zeh A. A prospective randomized radiographic and dual-energy X-ray absorptiometric study of migration and bone remodeling after implantation of two modern short-stemmed femoral prostheses. J Orthop Traumatol 2015;16:237243.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 42.

    Ercan A, Sokkar SM, Schmid G, Filler TJ, Abdelkafy A, Jerosch J. Periprosthetic bone density changes after MiniHip TM cementless femoral short stem: one-year results of dual-energy X-ray absorptiometry study. SICOT J 2016;2:40.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 43.

    Jahnke A, Engl S & Altmeyer Cet al. Changes of periprosthetic bone density after a cementless short hip stem: a clinical and radiological analysis. Int Orthop 2014;38:20452050.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 44.

    Lerch M, von der Haar-Tran A, Windhagen H, Behrens BA, Wefstaedt P, Stukenborg-Colsman CM. Bone remodelling around the Metha short stem in total hip arthroplasty: a prospective dual-energy X-ray absorptiometry study. Int Orthop 2012;36:533538.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 45.

    Shafy TA, Sayed A, Abdelazeem AH. Study of the bone behavior around a neck preserving short stem implant: bone densitometric analysis over a span of two years. SICOT J 2016;2:31.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 46.

    Steens W, Boettner F, Bader R, Skripitz R, Schneeberger A. Bone mineral density after implantation of a femoral neck hip prosthesis: a prospective 5 year follow-up. BMC Musculoskelet Disord 2015;16:192.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 47.

    Synder M, Krajewski K, Sibinski M, Drobniewski M. Periprosthetic bone remodeling around short stem. Orthopedics 2015;38:S40S45.

  • 48.

    Zeh A, Pankow F, Röllinhoff M, Delank S, Wohlrab D. A prospective dual-energy X-ray absorptiometry study of bone remodeling after implantation of the Nanos short-stemmed prosthesis. Acta Orthop Belg 2013;79:174180.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 49.

    Parchi PD, Ciapini G, Castellini I, Mannucci C, Nucci A-M, Piolanti N, Maffei S, Lisanti M. Evaluation of the effects of the Metha® short stem on periprosthetic bone remodelling in total hip arthroplasties: results at 48 months. Surg Technol Int 2017;30:346351.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 50.

    Boller S, Jahnke A, Augustin L, Ahmed G, Rickert M, Alexander B. Age-related osseointegration of a short hip stem: a clinical and radiological 24 months follow-up. Arch Orthop Trauma Surg 2019;139:405410.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 51.

    Nysted M, Benum P, Klaksvik J, Foss O, Aamodt A. Periprosthetic bone loss after insertion of an uncemented, customized femoral stem and an uncemented anatomical stem: a randomized DXA study with 5-year follow-up. Acta Orthop 2011;82:410416.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 52.

    Arabmotlagh M, Hennigs T, Rittmeister M. Femoral periprosthetic bone remodelling after implantation of custom made anatomic and standard straight stem hip prostheses. Z Orthop Ihre Grenzgeb 2003;141:519525.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 53.

    Cohen B, Rushton N. Accuracy of DEXA measurement of bone mineral density after total hip arthroplasty. J Bone Joint Surg [Br] 1995;77-B:77479.

  • 54.

    Grochola LF, Habermann B, Mastrodomenico N, Kurth A. Comparison of periprosthetic bone remodelling after implantation of anatomic and straight stem prostheses in total hip arthroplasty. Arch Orthop Trauma Surg 2008;128:383392.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 55.

    Boyle C, Kim IY. Comparison of different hip prosthesis shapes considering micro-level bone remodeling and stress-shielding criteria using three-dimensional design space topology optimization. J Biomech 2011;44:17221728.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 56.

    Brinkmann V, Radetzki F, Gutteck N, Delank S, Zeh A. Influence of varus/valgus positioning of the Nanos® and Metha® short-stemmed prostheses on stress shielding of metaphyseal bone. Acta Orthop Belg 2017;83:5766.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 57.

    Duncan CP, Masterson EL, Masri BA. Impaction allografting with cement for the management of femoral bone loss. Orthop Clin North Am 1998;29:297305.

  • 58.

    Goff TAJ, Bobak P. Femoral impaction allografting for significant bone loss in revision hip arthroplasty. Hip Int 2017;27:281285.

  • 59.

    Abdel MP, Roth PV, Harmsen WS, Berry DJ. What is the lifetime risk of revision for patients undergoing total hip arthroplasty? A 40-year observational study of patients treated with the Charnley cemented total hip arthroplasty. J Bone Joint Surg [Br] 2016;98-B:14361440.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 60.

    Cnudde PHJ, Nemes S, Bülow E, Timperley AJ, Whitehouse SL & Kärrholm Jet al. Risk of further surgery on the same or opposite side and mortality after primary total hip arthroplasty: a multi-state analysis of 133,654 patients from the Swedish Hip Arthroplasty Register. Acta Orthop 2018;89:386393.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 61.

    Kärrholm J, Mohaddes M, Odin D, Vinblad J, Rogmark C, Rolfson O. Swedish Hip Arthroplasty Register annual report 2017. Gothenburg: Swedish Hip Arthroplasty Register, 2018.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 62.

    DeRogatis MJ, Issack PS. Total femoral replacement as a salvage operation for the treatment of massive femoral bone loss during revision total hip arthroplasty. JBJS Rev 2018;6:e9.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 63.

    Sculco PK, Abdel MP, Lewallen DG. Management of femoral bone loss in revision total hip arthroplasty. Hip Int 2015;25:380387.

  • 64.

    Glassman AH, Bobyn JD, Tanzer M. New femoral designs: do they influence stress shielding? Clin Orthop Relat Res 2006;453:6474.

  • 65.

    Sheth NP, Melnic CM, Rozell JC, Paprosky WG. Management of severe femoral bone loss in revision total hip arthroplasty. Orthop Clin North Am 2015;46:329342, ix.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 66.

    Götze C, Ehrenbrink J, Ehrenbrink H. Bleibt der Krafteinfluss der Kurzschaftprothese auf den methaphysären proximalen Femur begrenzt? Osteodensitometrische Analysen der NANOS®-Schaftendoprothese. Z Orthop Unfall 2010;148:398405.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 67.

    Thorey F, Hoefer C, Abdi-Tabari N, Lerch M, Budde S, Windhagen H. Clinical results of the Metha short hip stem: a perspective for younger patients? Orthop Rev (Pavia) 2013;5:e34.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 68.

    Boskey AL, Coleman R. Critical reviews in oral biology and medicine. J Dent Res 2010;89:13331348.

  • 69.

    Venesmaa PK, Ger HPJK, Miettinen HJA, Jurvelin JS, Suomalainen OT, Alhava EM. Monitoring of periprosthetic BMD after uncemented total hip arthroplasty with dual-energy X-ray absorptiometry: a 3-year follow-up study. J Bone Miner Res 2001;16:10561061.

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 70.

    Hayashi S, Hashimoto S, Kanzaki N, Kuroda R, Kurosaka M. Stem anteversion affects periprosthetic bone mineral density after total hip arthroplasty. Hip Int 2016;26:260264.

    • PubMed
    • Search Google Scholar
    • Export Citation