Taper corrosion: a complication of total hip arthroplasty

in EFORT Open Reviews
Authors:
Michael M. Morlock TUHH Hamburg University of Technology, Hamburg, Germany.

Search for other papers by Michael M. Morlock in
Current site
Google Scholar
PubMed
Close
,
Robert Hube OCM Clinic Munich, Munich, Germany

Search for other papers by Robert Hube in
Current site
Google Scholar
PubMed
Close
,
Georgi Wassilew Department for Orthopaedics and Orthopaedic Surgery, University of Greifswald, Greifswald, Germany

Search for other papers by Georgi Wassilew in
Current site
Google Scholar
PubMed
Close
,
Felix Prange TUHH Hamburg University of Technology, Hamburg, Germany.

Search for other papers by Felix Prange in
Current site
Google Scholar
PubMed
Close
,
Gerd Huber TUHH Hamburg University of Technology, Hamburg, Germany.

Search for other papers by Gerd Huber in
Current site
Google Scholar
PubMed
Close
, and
Carsten Perka Center for Musculoskeletal Surgery, Orthopedic Department, Charité – Universitätsmedizin Berlin, Berlin, Germany

Search for other papers by Carsten Perka in
Current site
Google Scholar
PubMed
Close

Michael M. Morlock, Institute of Biomechanics, TUHH Hamburg University of Technology, Denickestrasse 15, 21073 Hamburg, Germany. Email: morlock@tuhh.de
Open access

  • The focus on taper corrosion in modular hip arthroplasty increased around 2007 as a result of clinical problems with large-head metal-on-metal (MoM) bearings on standard stems. Corrosion problems with bi-modular primary hip stems focused attention on this issue even more.

  • Factors increasing the risk of taper corrosion were identified in laboratory and retrieval studies: stiffness of the stem neck, taper diameter and design, head diameter, offset, assembly force, head and stem material and loading.

  • The high variability of the occurrence of corrosion in the clinical application highlights its multi-factorial nature, identifying the implantation procedure and patient-related factors as important additional factors for taper corrosion.

  • Discontinuing the use of MoM has reduced the revisions due to metal-related pathologies dramatically from 49.7% (MoM > 32 mm), over 9.2% (MoM ⩽ 32 mm) to 0.8% (excluding all MoM).

  • Further reduction can be achieved by omitting less stiff Ti-alloys and large metal heads (36 mm and above) against polyethylene (PE).

  • Standardized taper assembly of smaller and ceramic heads will reduce the clinical occurrence of taper corrosion even further. If 36 mm heads are clinically indicated, only ceramic heads should be used.

  • Taper-related problems will not comprise a major clinical problem anymore if the mentioned factors are respected.

Cite this article: EFORT Open Rev 2020;5:776-784. DOI: 10.1302/2058-5241.5.200013

Abstract

  • The focus on taper corrosion in modular hip arthroplasty increased around 2007 as a result of clinical problems with large-head metal-on-metal (MoM) bearings on standard stems. Corrosion problems with bi-modular primary hip stems focused attention on this issue even more.

  • Factors increasing the risk of taper corrosion were identified in laboratory and retrieval studies: stiffness of the stem neck, taper diameter and design, head diameter, offset, assembly force, head and stem material and loading.

  • The high variability of the occurrence of corrosion in the clinical application highlights its multi-factorial nature, identifying the implantation procedure and patient-related factors as important additional factors for taper corrosion.

  • Discontinuing the use of MoM has reduced the revisions due to metal-related pathologies dramatically from 49.7% (MoM > 32 mm), over 9.2% (MoM ⩽ 32 mm) to 0.8% (excluding all MoM).

  • Further reduction can be achieved by omitting less stiff Ti-alloys and large metal heads (36 mm and above) against polyethylene (PE).

  • Standardized taper assembly of smaller and ceramic heads will reduce the clinical occurrence of taper corrosion even further. If 36 mm heads are clinically indicated, only ceramic heads should be used.

  • Taper-related problems will not comprise a major clinical problem anymore if the mentioned factors are respected.

Cite this article: EFORT Open Rev 2020;5:776-784. DOI: 10.1302/2058-5241.5.200013

Corrosion at modular taper junctions has been a known phenomenon for the last 30 years. Interestingly the first paper addressing the issue came to the conclusion that ‘there should be no occurrence of long-term in-vivo fretting corrosion’ of heads made from Cobald-Chromium-Alloy or Ti6Al4V stems. 1 Today it is well known that such a general statement was not justified based on pre-clinical testing alone, which is always a drastic simplification of in-vivo loading and environment. A PubMed literature search (13 July 2020) with the search terms ‘(taper OR trunnion) AND (hip) AND (corrosion)’ yielded 350 relevant hits, demonstrating the continuing interest in this topic. The reported studies describe clinical observations based on explant analyses from single cases or case series or the results of laboratory testing, systematically investigating the influence of the parameters derived from the clinical observations on the risk of taper corrosion. The main findings of these reports are summarized in this review, but the goal is not to add another review article to this topic but rather to focus on the question: How big is the taper corrosion problem currently in a representative setting in Germany after the removal of the most endangered designs from the market? The second goal is to formulate measures based on this analysis to further reduce the problem in the future.

History

Design

The original Morse taper was invented by Stephen Morse in the 1860s for machine tool operators, allowing them to install or remove tool bits quick and easily, and is still used in contemporary machines. It had a relatively small angle of 2.9° resulting in a steep taper angle. 2 The Morse taper was introduced in Total Hip Arthroplasty (THA) at the end of the 1970s for the assembly of ceramic heads to metal stems after problems with the fixation strength of ceramic heads fixed by glueing or screwing. 3 They became widely popular around 1985 to allow the combination of different materials in general, i.e. also cobalt–chromium heads on titanium-alloy stems, and to adjust combined offset and leg length intraoperatively. The original taper dimension in THA was ‘14/16’, which refers to a truncated cone of 20 mm length with diameters of 14 mm and 16 mm at either end (resulting in the abovementioned taper angle of 5.725°). This taper angle had been shown to be favourable for higher fracture loads of ceramic heads. 4 In the 1990s, the taper dimensions were reduced to smaller diameters such as 12/14, 11/13 or even 9/10 and new denotations such as C-taper, Type-1 taper or V40 taper were introduced. A reduction in taper length occurred concomitantly (from 20 mm to below 10 mm). 5 All these changes were made to increase the possible range of motion of the joint with the use of small ball heads (32 mm diameter and below). The downside of all these changes was a reduction of taper contact area and taper bending stiffness. Today in Europe the so-called ‘12/14 taper’ or ‘Euro taper’ is the most frequently used design (with a similar angle as the original ‘14/16 taper’, Fig. 1). There are some important basic facts about metal tapers, which help to understand the taper problems observed:

  • (1) There is not now and never has been a standard defining specific taper dimension. As a consequence, tapers between (and even within) manufacturers vary. Even so, they might be named similarly.

  • (2) Tapers were designed for and perform best under torsional loading around the taper axis and poorly under bending loading (as in high offset stems, long heads, large heads, and especially for modular necks and revision stems).

Fig. 1
Fig. 1

Definition of the head-stem taper connection. The male stem taper geometry is defined by the proximal and distal diameters and its length. The ‘12/14’ taper is defined with a proximal diameter of 12 mm, a distal diameter of 14 mm and a length of 20 mm, resulting in a male stem taper angle of 5.725° or 5°43’30’’ (the same angle as in the original ‘14/16’ taper). The female taper angle of ceramic heads is always by 1’ to 6’ larger, in order to achieve initial proximal engagement between female and male taper. 6 Most metal heads are designed similarly but some are manufactured with the same nominal female taper angle as the stem. The true contact length depends on the assembly force and the taper angle difference (indicated in green). 85 All these values are not standardized and vary between manufacturers. 86

Citation: EFORT Open Reviews 5, 11; 10.1302/2058-5241.5.200013

Clinical situation – Metal-on-metal bearings

Since dislocation was one of the most frequent revision reasons, larger heads were used more frequently, starting around 2004, to reduce the dislocation risk. 6 Since larger heads generate more wear against polyethylene (PE), metal-on-metal (MoM) hard-on-hard articulations started to become popular to enable larger head sizes. Opposite to PE bearings, the amount of wear in MoM bearings decreases with head size – as long as they are lubricated properly. This was supported by good results for hip resurfacing arthroplasty. 7 In unfavourable lubrication situations, however, MoM bearings can generate high friction, which in turn increases the bending and torsion torque at the taper interface.

Clinically, large MoM bearings on standard hip stems did not perform well from the beginning. In 2009, a prospective study comparing hip resurfacing to large MoM bearings for a similar design caused great concern, since the metal ion concentrations in the MoM group were found to be elevated. 8 Since the taper junction was the only difference between the two groups in this study, the taper problem was ‘identified’. Prior to this publication, the taper interface was not seen as a critical factor since the focus in MoM was solely on wear and friction of these bearings. As a consequence, no distinction between hip resurfacing and large MoM was made. 911 In the 2009 annual report of the National Joint Registries of England and Wales (NJR), the results of resurfacing and large-head MoM are consequently reported together. 12 In 2010, when the problems with large-head MoM prostheses were starting to become evident, the results were reported separately. 6 The scientific interest in this topic, as reflected in the number of publications, started to rise right around this time (Fig. 2).

Fig. 2
Fig. 2

Number of peer-reviewed journal publications for the search terms ‘(taper OR trunnion) AND hip AND corrosion’ in PubMed. The number of publications started to increase from the few anecdotal reports published before 2009 and peaked in 2016 with continuing high interest since.

Citation: EFORT Open Reviews 5, 11; 10.1302/2058-5241.5.200013

Initially one implant design was mainly held responsible for the taper corrosion problems observed (the Articular Surface Replacement (ASR) design by DePuy). 13,14 This device was recalled in 2010, which, however, did not resolve the problem. It quickly became clear that the whole ‘category’ of large-head MoM exhibited a much higher risk of taper corrosion in comparison with smaller bearings against PE. The great public interest in the corrosion issue is due to the fact that the metal debris produced by wear or corrosion can be directly associated with clinical symptoms leading to revision, which was first reported for resurfacing 15,16 and soon after for large MoM bearings. 17,18 In MoM bearings the material loss from the cup bearing surface and the female taper were shown to demonstrate the greatest material loss. 19

Stem taper fracture as the worst clinical endpoint of taper corrosion has been observed in only one particular MoM design using a Ti-alloy adapter sleeve connector with a Ti-alloy stem (Fig. 3). In this design a high frequency of cold welding between stem and adapter was also observed. 20

Fig. 3
Fig. 3

(a) X-ray of a stem taper fracture in a Ti-alloy primary stem with a large MoM bearing after four years in situ. This design utilizes a Ti-alloy adapter sleeve to assemble the CoCr-bearing surface to the stem. (b) The explanted components with the broken stem taper inside the adapter sleeve inside the head (courtesy H. Ettema, Isala).

Citation: EFORT Open Reviews 5, 11; 10.1302/2058-5241.5.200013

The risk of taper corrosion in large MoM THAs was ultimately taken up politically by the European Commission with the establishment of a Scientific Committee on Emerging and Newly Identified Health Risks (SCENIHR) working group investigating ‘The safety of Metal-on-Metal joint replacements with a particular focus on hip implants’. The consensus of this working group, advising against the use of large MoM bearings, was implemented shortly afterwards. 21 Discontinuing the use of MoM has reduced the revisions due to metal-related pathologies dramatically from 49.7% (MoM > 32 mm), over 9.2% (MoM ⩽ 32 mm) to 0.8% (excluding all MoM). 22,23 . However, the discontinuation of large MoM bearings did not completely eliminate the taper corrosion issue.

Clinical situation – bearings utilizing PE

Just around the time that MoM THA was discontinued, several reports reporting taper corrosion with metal heads against PE were published and this continues today. 24,25 Catastrophic gross stem taper failures (GTF) due to corrosion and abrasive taper wear are the newest addition to the taper corrosion story. 26 Such failures are reported mostly for one particular stem design with a small V-40 taper made from a less stiff Ti-12Mo-6Zr-2Fe titanium-alloy (TMZF; Accolade I, Stryker, Mahwah, New Jersey, USA) in combination with heads made of Cobald-Chromium-Alloy (CoCr) of sizes of mostly 36 mm and above. 2731

Factors influencing the occurrence of taper corrosion

Any modular connection involving metal alloys will show some corrosion if in contact with body fluids and exposed to micromotion. 3234 Tapers were introduced in engineering for the transfer of loads applied along the taper axis. Loads applied off-axis – as is the normal situation for the head-stem taper connection in THA – together with joint friction moments, cause a bending load at the taper interface with non-symmetrical compressive radial (‘hoop’) stress distributions. This stress asymmetry results in micromotions or even a loss of contact during cyclic loading, leading to fretting and crevice corrosion (Fig. 4). The magnitude of micromotion at the taper interface is influenced by several factors, which can be categorized into three groups:

  • - design and material;

  • - assembly (surgeon factors);

  • - loading (patient factors).

Fig. 4
Fig. 4

(a) Female head taper corrosion after seven years in situ. The overlay shows the depth of the female head taper wear (material loss 6.7 mm³). (b) The male stem taper did not show signs of corrosion but slight marks from the revision surgery.

Citation: EFORT Open Reviews 5, 11; 10.1302/2058-5241.5.200013

Design and material

Taper surface morphology (rough or smooth, long or short) has been shown to influence the amount of corrosion observed but the data are still insufficient to draw definite conclusions. 3537 Even so it was shown early on that low neck stiffness is a contributing factor for corrosion, 32 changes in taper design over the last 20 years have led to even more flexible tapers, 38 which consequently show more corrosion. 39

Long and large heads, varus stems and higher offset stems apply higher load to the taper interface and as such exhibit more fretting and corrosion. 4047 Fracture of hip revision stems are a ‘good’ example to illustrate this (Fig. 5): Due to the long lever arm of the joint load to the stem–neck piece taper connection, the risk of micro-movement with consecutive corrosion is clearly elevated, especially under unfavourable loading conditions, ultimately resulting in an elevated rate of taper fracture at the neck piece stem taper. 48 Similar arguments can explain the higher corrosion problems for dual taper hips. 49,50

Fig. 5
Fig. 5

(a) X-ray after fracture of of a Ti-Ti Revision stem after seven years in situ (courtesy M. Rudert, Würzburg). (b) Fracture surfaces of the broken taper showing the typical split lines of a fatigue fracture.

Citation: EFORT Open Reviews 5, 11; 10.1302/2058-5241.5.200013

Finally, the frequency of metal debris-related revisions was shown to be greatly reduced when using ceramic heads, 5153 especially in the combination with PE as an acetabular bearing partner. 54,55

In modular hemiarthroplasty, taper corrosion seems to play a much smaller role, however, a few cases have been reported. 5659

Alloy combinations

Corrosion at Ti-alloy–Ti-alloy taper junctions is typically not recognized until catastrophic failure occurs as a result of the absence of a biological response to the Ti debris in most cases (i.e. fracture, Fig. 3, Fig. 5). In contrast, clinical failure of taper junctions between CoCr heads and Ti-alloy stems typically does not involve mechanical failure but is caused by the biological response to the Co and Cr debris, which is created in the corrosion process (Fig. 4). 34 It is important to realize that the male stem taper is usually not damaged in such an situation, even so it might look corroded due to the Co and Cr debris from the inside head taper. After careful cleaning it is possible in most cases to maintain the stem and put on a ceramic head with a Ti-alloy adapter sleeve. 54 Both CoCr heads and ceramic heads on CoCr stems show a very low material release. 60 The results for oxidized zirconium heads in comparison to CoCr heads are favourable but inconsistent. 51,61

Fig. 6
Fig. 6

(a) X-ray of a male patient (76 years) prior the 4th isolated right cup revision in 2019 after original stem implantation with a 32 mm metal XXXL head in 1984. (b) The stem taper showed only light corrosion and minor damage from the multiple revision surgeries. (c) X-ray after the revision with a 36 mm ceramic option XXXL head on the original stem.

Citation: EFORT Open Reviews 5, 11; 10.1302/2058-5241.5.200013

Assembly

The assembly condition plays an important role in the strength of the taper connection. 62,63 Minimal invasive surgery with small incisions might indirectly also be associated with the frequency of taper corrosion due to the challenges in cleaning the taper and to apply appropriate assembly forces in taper direction. Tapers are designed for clean and dry assembly conditions. Assembling them in a contaminated situation (blood, bone, water, fat) causes an increase in micromotion during loading or a reduction in fracture strength of ceramic heads. 6368 Contamination might also reduce secondary seating after implantation due to the component of the joint contact force directed into the direction of the taper axis. 69 This is the reason why stem tapers should always be carefully cleaned, rinsed and dried before assembly. Insufficient assembly force causes less firm fixation, with a risk of increased micromotion during loading. 7072 Contact area and fixation strength increase linearly with impaction force. 71,73 Larger heads require even higher assembly forces since their higher joint friction moments must be withstood by the taper interface. 74

Loading

The loading magnitude and direction directly influences the micromotion at the taper interface. 75 A continuous increase in the obesity rates in most populations, reaching 42.4% in the US in 2017 to 2018, 76 has continuously increased the load on the taper interface, 77 especially in combination with larger head sizes, which is a common trend, especially in the US. 78 In the last few years a further aspect has been highlighted: the composition of the body fluids surrounding the taper junction. It could be shown that inflammatory conditions enhance the risk of corrosion of CoCr and Ti-alloys. 79,80

Despite the influence of the mentioned parameters on the loading of the taper, corrosion has been shown for nearly all taper designs and implant configurations. 81

Present situation in Germany

In order to get an impression of the present magnitude of the taper corrosion problem in Germany, an observational study in cooperation with four of the leading German orthopaedic hospitals was performed (Charite Berlin, Endo-Klinik Hamburg, OCM Munich, Uni Greifswald). Intra-operative pictures of the male stem taper during revision surgeries, in planned cup revision surgeries, in which the stem was expected to be maintained, were taken. These qualitative pictures, in combination with the revision reason specified in the surgical report, were used for the classification of the revision reason in 100 cases. Not all revision cases in the respective hospitals were documented due to time limitations in the surgical routine. As a consequence, a rather arbitrary and random selection was achieved. Head taper corrosion was the reason for revision in 3% of the cases: one case with corrosion of the female head taper in a large MoM bearing (50 mm) and two cases with corrosion of the female head taper in 36 mm metal heads against PE (Fig. 4). In one case of corrosion at the neck taper of a bi-modular stem design, the stem had to be removed (1% of the cases). This confirms the general perception in Germany that taper corrosion problems predominantly occur for designs already identified at risk, which were not frequently used in Germany in the past. This is also reflected by the fact that metal debris-related revision is not even listed as a revision reason in the Endoprothesen Register Deutschland (EPRD) German Arthroplasty Report 2019. 82

How to minimize future taper problems in the patient

THA is frequently referred to as the ‘Operation of the 20th century’ and modular implant systems have contributed greatly to this success. 83 In this review, the history of the evolution of taper design together with the evidence on the factors, which are thought to influence the frequency of clinically observed taper corrosion, are summarized. It is impressive how much research has been conducted and still is conducted on this issue (Fig. 2). There is wide agreement that taper corrosion is a multi-factorial phenomenon influenced by the three factor groups addressed in this review: mechanical loading (head size, offset, friction, patient weight and activity, geometrical configuration), the reduction of loading capacity due to recent design changes (taper length, taper diameter, softer alloys, taper surface morphology) and the rather underemphasized importance of contamination and proper assembly. Furthermore, corrosion of metal alloys always occurs in the human body due to the aggressive environment, but does not necessarily become a clinical problem as long as the amount of corrosion is small. 34,42,84

The final and most important question concerns the further reduction of the frequency of taper corrosion problems. Even so, not all aspects of the corrosion mechanism in-vivo are known yet. There is more than enough evidence that respecting only one of the decisive factors alone will not solve the taper corrosion problem but that all factors have to be addressed simultaneously. The suggestions given below summarize the required measures:

  1. The use of 36 mm metal heads should only be for a good clinical reason and if it can be ascertained that they are assembled properly to a clean taper. The safer option is the use of a 36 mm ceramic head.

  2. A reduction in head size reduces the occurrence of taper corrosion.

  3. Stems made from less stiff metal alloys in combination with small tapers should be omitted. Small tapers, especially in combination with large heads, should be used carefully.

  4. Large offsets, especially in combination with long heads, carry a higher risk for taper corrosion.

  5. The use of ceramic heads reduces the taper corrosion risk significantly.

  6. Careful and proper cleaning and assembly reduces the risk of taper corrosion in all configurations.

Taper corrosion is a complication that can occur in any THA, independent of design and head size, as long as stems are made from metallic alloys. 81 Respecting the mentioned measures will further reduce the magnitude of the problem, which was already greatly reduced by the elimination of MoM THA bearings. The very low observed rate of taper corrosion in Germany is probably mostly attributed to the long-term education of the proper use of ceramic heads (‘clean and impact’) and the traditional use of smaller head diameters in metal heads: 54 In 2019, 88% of all implanted heads were ceramic and the dominant head size was 32 mm (54%). 82 If properly assembled and not exposed to excessive loading, metal heads can also provide very good long-time survival in the patient, even in adverse loading conditions if assembled properly (Fig. 6), but the diameter of the heads should not exceed 32 mm.

Open access

This article is distributed under the terms of the Creative Commons Attribution-Non Commercial 4.0 International (CC BY-NC 4.0) licence (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed.

The observational clinical study was financially supported by Ceramtec.

ICMJE Conflict of interest statement

MMM reports that Ceramtec supported the clinical observational study financially as part of a larger taper damage study. He reports consultancy for DePuy Synthes, providing expert testimony for Zimmer-Biomet, grants/grants pending with DePuy Synthes, and payment for lectures including service on speakers’ bureaus from DePuy Synthes, Implantcast and Lima, all outside the submitted work.

RH reports board membership of Zimmer-Biomet’s Resident Education Program, payment for lectures including service on speakers’ bureaus from Zimmer-Biomet and royalties from Zimmer-Biomet, all outside the submitted work.

GH reports a grant from Ceramtec, related to the submitted work, and grants/grants pending from DePuy Synthes outside the submitted work.

CP reports board membership of Bone and Joint Journal (travel and accommodation only), consultancy for DePuy/Synthes, LINK and Zimmer, payment for lectures including service on speakers’ bureaus from LINK, royalties from DePuy/Synthes, Smith & Nephew and Zimmer and payment for development of educational presentations from AORecon, all outside the submitted work.

The other authors declare no conflict of interest relevant to this work.

Funding statement

The author or one or more of the authors have received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this article.

OA licence text

This article is distributed under the terms of the Creative Commons Attribution-Non Commercial 4.0 International (CC BY-NC 4.0) licence (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed.

References

  • 1.

    Fricker DC , Shivanath R . Fretting corrosion studies of universal femoral head prostheses and cone taper spigots. Biomaterials 1990 Sep;11:495500 .

  • 2.

    Hussenbocus S , Kosuge D , Solomon LB , Howie DW , Oskouei RH . Head-neck taper corrosion in hip arthroplasty. Biomed Res Int 2015:758123 .

  • 3.

    Piconi C , Streicher RM . Forty years of ceramic-on-ceramic THR bearings. Semin Arthroplasty 2013; 24:188192 .

  • 4.

    Willmann G . Das prinzip der konus-steckverbindung für keramische kugelköpfe bei hüftendendoprothesen. J Mater Sci Technol 1993; 24:315319 .

  • 5.

    Werner PH , Ettema HB , Witt F , Morlock MM , Verheyen CC . Basic principles and uniform terminology for the head-neck junction in hip replacement. Hip Int 2015; 25:115119 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 6.

    National Joint Registry. 7th Annual Report for England and Wales. 2010 .

  • 7.

    McMinn D , Daniel J . History and modern concepts in surface replacement. Proc Inst Mech Eng H 2006; 220:239251. H .

  • 8.

    Garbuz DS , Tanzer M , Greidanus NV , Masri BA , Duncan CP . The John Charnley Award: Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res 2010; 468:318325 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 9.

    Nassutt R , Wimmer MA , Schneider E , Morlock MM . The influence of resting periods on friction in the artificial hip. Clin Orthop Relat Res 2003:127138 .

  • 10.

    Morlock MM , Bishop N , Rüther W , Delling G , Hahn M . Biomechanical, morphological, and histological analysis of early failures in hip resurfacing arthrosplasty. Proc Inst Mech Eng H 2006; 220:333344 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 11.

    Morlock MM , Bishop N & Stahmer F et al. Reasons for failure of hip resurfacing implants. A failure analysis based on 250 revision specimens. Orthopade 2008; 37:695703 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 12.

    National Joint Registry. 6th Annual Report for England and Wales. 2009 .

  • 13.

    Langton DJ , Jameson SS , Joyce TJ , Hallab NJ , Natu S , Nargol AV . Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear. J Bone Joint Surg Br 2010; 92:3846 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 14.

    Langton DJ , Jameson SS & Joyce TJ et al. Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br 2011; 93:10111016 .

  • 15.

    Pandit H , Vlychou M & Whitwell D et al. Necrotic granulomatous pseudotumours in bilateral resurfacing hip arthoplasties: evidence for a type IV immune response. Virchows Arch 2008; 453:529534 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 16.

    Pandit H , Glyn-Jones S & McLardy-Smith P et al. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br 2008; 90:847851 .

  • 17.

    Crawford R , Ranawat CS , Rothman RH . Metal on metal: is it worth the risk? J Arthroplasty 2010; 25:12 .

  • 18.

    Hothi HS , Berber R , Whittaker RK , Blunn GW , Skinner JA , Hart AJ . The relationship between cobalt/chromium ratios and the high prevalence of head-stem junction corrosion in metal-on-metal total hip arthroplasty. J Arthroplasty 2016; 31:11231127 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 19.

    Matthies AK , Racasan R & Bills P et al. Material loss at the taper junction of retrieved large head metal-on-metal total hip replacements. J Orthop Res 2013; 31:16771685 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 20.

    Whittaker RK , Zaghloul AM & Hothi HS et al. Clinical cold welding of the modular total hip arthroplasty prosthesis. J Arthroplasty 2017; 32:610615 .

  • 21.

    Hannemann F , Hartmann A & Schmitt J et al. European multidisciplinary consensus statement on the use and monitoring of metal-on-metal bearings for total hip replacement and hip resurfacing. Orthop Traumatol Surg Res 2013; 99:263271 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 22.

    Australian Orthopedic Association. Annual Report Hip and Knee Arthroplasty of the National Joint Registry 2019. aoanjrr.dmac.adelaide.edu.au/annual-reports-2019; 2019 (date last accessed 28 September 2020).

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 23.

    Australian Orthopaedic Association National Joint Replacement Registry. Metal/Metal Bearing Surface in Total Conventional Hip Arthroplasty 2019. aoanjrr.dmac.adelaide.edu.au/annual-reports-2019; 2019 (date last accessed 28 September 2020).

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 24.

    Cooper HJ , Della Valle CJ & Berger RA et al. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am 2012; 94:16551661 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 25.

    McGrory BJ , MacKenzie J , Babikian G . A high prevalence of corrosion at the head-neck taper with contemporary zimmer non-cemented femoral hip components. J Arthroplasty 2015; 30:12651268 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 26.

    Bansal T , Aggarwal S , Dhillon MS , Patel S . Gross trunnion failure in metal on polyethylene total hip arthroplasty–a systematic review of literature. Int Orthop 2020; 44:609621 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 27.

    Morlock MM , Dickinson EC , Günther KP , Bünte D , Polster V . Head taper corrosion causing head bottoming out and consecutive gross stem taper failure in total hip arthroplasty. J Arthroplasty 2018; 33:35813590 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 28.

    Fahey EJ , Pomeroy E , Rowan FE . Acute trunnion failure of a TMZF alloy stem with large diameter femoral heads. J Orthop 2020; 20:1720 .

  • 29.

    Matsen Ko L , Chen AF , Deirmengian GK , Hozack WJ , Sharkey PF . Catastrophic femoral head-stem trunnion dissociation secondary to corrosion. J Bone Joint Surg Am 2016; 98:14001404 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 30.

    Banerjee S , Cherian JJ & Bono JV et al. Gross trunnion failure after primary total hip arthroplasty. J Arthroplasty 2015; 30:641648 .

  • 31.

    Walker PS , Campbell PA , Brazil D , McTighe T . Trunnion corrosion and early failure in monolithic metal-on-polyethylene tmzf femoral components: A case series. Reconstr Rev 2016; 6:1318 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 32.

    Goldberg JR , Gilbert JL , Jacobs JJ , Bauer TW , Paprosky W , Leurgans S . A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res 2002:149161 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 33.

    Gilbert JL , Buckley CA , Jacobs JJ . In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res 1993; 27:15331544 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 34.

    Lutzner J , Gunther KP , Postler A , Morlock M . Metal ion release after hip and knee arthroplasty - causes, biological effects and diagnostics. Z Orthop Unfall 2020 Aug;158:369382 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 35.

    Panagiotidou A , Meswania J , Hua J , Muirhead-Allwood S , Hart A , Blunn G . Enhanced wear and corrosion in modular tapers in total hip replacement is associated with the contact area and surface topography. J Orthop Res 2013; 31:20322039 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 36.

    Whittaker RK , Hothi HS , Eskelinen A , Blunn GW , Skinner JA , Hart AJ . Variation in taper surface roughness for a single design effects the wear rate in total hip arthroplasty. J Orthop Res 2017; 35:17841792 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 37.

    Hothi HS , Whittaker RK , Meswania JM , Blunn GW , Skinner JA , Hart AJ . Influence of stem type on material loss at the metal-on-metal pinnacle taper junction. Proc Inst Mech Eng H 2015; 229:9197 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 38.

    Porter DA , Urban RM , Jacobs JJ , Gilbert JL , Rodriguez JA , Cooper HJ . Modern trunnions are more flexible: a mechanical analysis of THA taper designs. Clin Orthop Relat Res 2014; 472:39633970 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 39.

    Tan SC , Teeter MG , Del Balso C , Howard JL , Lanting BA . Effect of taper design on trunnionosis in metal on polyethylene total hip arthroplasty. J Arthroplasty 2015; 30:12691272 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 40.

    Norman TL , Denen JE , Land AJ , Kienitz DM , Fehring TA . Taper-trunnion interface stress varies significantly with head size and activity. J Arthroplasty 2019; 34:157162 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 41.

    Del Balso C , Teeter MG , Tan SC , Lanting BA , Howard JL . Taperosis: does head length affect fretting and corrosion in total hip arthroplasty? Bone Joint J 2015; 97-B:911-916 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 42.

    Del Balso C , Teeter MG , Tan SC , Howard JL , Lanting BA . Trunnionosis: does head size affect fretting and corrosion in total hip arthroplasty? J Arthroplasty 2016; 31:23322336 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 43.

    Panagiotidou A , Meswania J & Osman K et al. The effect of frictional torque and bending moment on corrosion at the taper interface : an in vitro study. Bone Joint J 2015; 97-B:463472 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 44.

    Craig P , Bancroft G , Burton A , Collier S , Shaylor P , Sinha A . Raised levels of metal ions in the blood in patients who have undergone uncemented metal-on-polyethylene Trident-Accolade total hip replacement. Bone Joint J 2014; 96-B:4347 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 45.

    Lavernia CJ , Iacobelli DA , Villa JM , Jones K , Gonzalez JL , Jones WK . Trunnion-head stresses in tha: are big heads trouble? J Arthroplasty 2015; 30:10851088 .

  • 46.

    Hothi HS , Panagiotopoulos AC & Whittaker RK et al. Damage patterns at the head-stem taper junction helps understand the mechanisms of material loss. J Arthroplasty 2017; 32:291295 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 47.

    Hothi HS , Eskelinen AP & Berber R et al. Factors associated with trunnionosis in the metal-on-metal pinnacle hip. J Arthroplasty 2017; 32:286290 .

  • 48.

    Krull A , Morlock MM , Bishop NE . Factors influencing taper failure of modular revision hip stems. Med Eng Phys 2018; 54:6573 .

  • 49.

    Buente D , Huber G , Bishop N , Morlock M . Quantification of material loss from the neck piece taper junctions of a bimodular primary hip prosthesis. A retrieval study from 27 failed Rejuvenate bimodular hip arthroplasties. Bone Joint J 2015; 97-B:13501357 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 50.

    Di Laura A , Hothi HS , Henckel J , Kwon YM , Skinner JA , Hart AJ . Retrieval findings of recalled dual-taper hips. J Bone Joint Surg Am 2018; 100:16611672 .

  • 51.

    Tan SC , Lau AC , Del Balso C , Howard JL , Lanting BA , Teeter MG . Tribocorrosion: Ceramic and oxidized zirconium vs cobalt-chromium heads in total hip arthroplasty. J Arthroplasty 2016; 31:20642071 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 52.

    Kurtz SM , Kocagöz SB & Hanzlik JA et al. Do ceramic femoral heads reduce taper fretting corrosion in hip arthroplasty? A retrieval study. Clin Orthop Relat Res 2013; 471:32703282 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 53.

    White PB , Meftah M , Ranawat AS , Ranawat CS . A comparison of blood metal ions in total hip arthroplasty using metal and ceramic heads. J Arthroplasty 2016; 31:22152220 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 54.

    Matharu GS , Pandit HG , Murray DW , Judge A . Adverse reactions to metal debris occur with all types of hip replacement not just metal-on-metal hips: a retrospective observational study of 3340 revisions for adverse reactions to metal debris from the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man. BMC Musculoskelet Disord 2016; 17:495 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 55.

    Hothi HS , Eskelinen AP & Henckel J et al. Effect of bearing type on taper material loss in hips from 1 manufacturer. J Arthroplasty 2018; 33:15881593 .

  • 56.

    Del Balso C , Teeter MG , Tan SC , Lanting BA , Howard JL . Does the additional articulation in retrieved bipolar hemiarthroplasty implants decrease trunnionosis compared to total hip arthroplasty? J Arthroplasty 2018; 33:268272 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 57.

    Moore BF , Lachiewicz PF . Corrosion and adverse tissue reaction after modular unipolar hip hemiarthroplasty. Arthroplast Today 2017; 3:207210 .

  • 58.

    Whitehouse MR , Endo M , Masri BA . Adverse local tissue reaction associated with a modular hip hemiarthroplasty. Clin Orthop Relat Res 2013; 471:40824086 .

  • 59.

    Khair MM , Nam D , DiCarlo E , Su E . Aseptic lymphocyte dominated vasculitis-associated lesion resulting from trunnion corrosion in a cobalt-chrome unipolar hemiarthroplasty. J Arthroplasty 2013; 28:196.e11196.e14 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 60.

    Di Laura A , Hothi H & Henckel J et al. Retrieval analysis of metal and ceramic femoral heads on a single CoCr stem design. Bone Joint Res 2017; 6:345350 .

  • 61.

    Hampton C , Weitzler L , Baral E , Wright TM , Bostrom MPG . Do oxidized zirconium heads decrease tribocorrosion in total hip arthroplasty? A study of retrieved components. Bone Joint J 2019; 101-B:386389 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 62.

    Danoff JR , Longaray J , Rajaravivarma R , Gopalakrishnan A , Chen AF , Hozack WJ . Impaction force influences taper-trunnion stability in total hip arthroplasty. J Arthroplasty 2018; 33:S270S274 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 63.

    Krull A , Morlock MM , Bishop NE . The influence of contamination and cleaning on the strength of modular head taper fixation in total hip arthroplasty. J Arthroplasty 2017; 32:32003205 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 64.

    Lavernia CJ , Baerga L & Barrack RL et al. The effects of blood and fat on Morse taper disassembly forces. Am J Orthop (Belle Mead NJ) 2009; 38:187190 .

  • 65.

    Wuttke V , Witte H , Kempf K , Oberbach T , Delfosse D . Influence of various types of damage on the fracture strength of ceramic femoral heads. Biomed Tech (Berl) 2011; 56:333339 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 66.

    Morlock MM . The taper disaster—how could it happen? Hip Int 2015; 25:339346 .

  • 67.

    Jauch SY , Huber G , Hoenig E , Baxmann M , Grupp TM , Morlock MM . Influence of material coupling and assembly condition on the magnitude of micromotion at the stem-neck interface of a modular hip endoprosthesis. J Biomech 2011; 44:17471751 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 68.

    Schäfer R , Soltész U , Bernard PF . Friction in hip-joint prostheses and its influence on the fixation of the artificial head. J Mater Sci Mater Med 1998; 9:687690 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 69.

    Jauch SY , Coles LG , Ng LV , Miles AW , Gill HS . Low torque levels can initiate a removal of the passivation layer and cause fretting in modular hip stems. Med Eng Phys 2014; 36:11401146 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 70.

    Haschke H , Jauch-Matt SY , Sellenschloh K , Huber G , Morlock MM . Assembly force and taper angle difference influence the relative motion at the stem-neck interface of bi-modular hip prostheses. Proc Inst Mech Eng H 2016; 230:690699 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 71.

    Rehmer A , Bishop NE , Morlock MM . Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses. Clin Biomech 2012; 27:7783 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 72.

    Pennock AT , Schmidt AH , Bourgeault CA . Morse-type tapers: factors that may influence taper strength during total hip arthroplasty. J Arthroplasty 2002; 17:773778 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 73.

    Witt F , Gührs J , Morlock MM , Bishop NE . Quantification of the contact area at the head-stem taper interface of modular hip prostheses. PLoS One 2015; 10:e0135517 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 74.

    Bishop NE , Waldow F , Morlock MM . Friction moments of large metal-on-metal hip joint bearings and other modern designs. Med Eng Phys 2008; 30:10571064 .

  • 75.

    Jauch SY , Huber G & Sellenschloh K et al. Micromotions at the taper interface between stem and neck adapter of a bimodular hip prosthesis during activities of daily living. J Orthop Res 2013; 31:11651171 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 76.

    Hales C , Carroll M , Fryar C , Ogden C . Prevalence of obesity and severe obesity among adults: United States, 2017–2018. NCHS Data Brief 2020;(No. 360, Feb):18 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 77.

    Donaldson FE , Coburn JC , Siegel KL . Total hip arthroplasty head-neck contact mechanics: a stochastic investigation of key parameters. J Biomech 2014; 47:16341641 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 78.

    American Academy of Orthopaedic Surgeons (AAOS). 2019 Annual Report of the American Joint Replacement Registry (AJRR). Rosemont, IL: 2019 .

  • 79.

    Kubacki GW , Gilbert JL . The effect of the inflammatory species hypochlorous acid on the corrosion and surface damage of Ti-6Al-4V and CoCrMo alloys. J Biomed Mater Res A 2018; 106:31853194 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 80.

    Liu Y , Gilbert JL . The effect of simulated inflammatory conditions and Fenton chemistry on the electrochemistry of CoCrMo alloy. J Biomed Mater Res B Appl Biomater 2018; 106:209220 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 81.

    Jennings JM , Dennis DA , Yang CC . Corrosion of the head-neck junction after total hip arthroplasty. J Am Acad Orthop Surg 2016; 24:349356 .

  • 82.

    Endoprothesenregister Deutschland (EPRD). Jahresbericht 2019. Berlin: 2019 .

  • 83.

    Learmonth ID , Young C , Rorabeck C . The operation of the century: total hip replacement. Lancet 2007; 370:15081519 .

  • 84.

    Osman K , Panagiotidou AP , Khan M , Blunn G , Haddad FS . Corrosion at the head-neck interface of current designs of modular femoral components: essential questions and answers relating to corrosion in modular head-neck junctions. Bone Joint J 2016; 98–B:579584 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 85.

    Witt F , Gührs J , Morlock MM , Bishop NE . Quantification of the contact area at the head-stem taper interface of modular hip prostheses. PLoS One 2015; 10:e0135517 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 86.

    Mueller U , Braun S , Schroeder S , Sonntag R , Kretzer JP . Same same but different? 12/14 stem and head tapers in total hip arthroplasty. J Arthroplasty 2017; 32:31913199 .

    • PubMed
    • Search Google Scholar
    • Export Citation

 

  • Collapse
  • Expand
  • Fig. 1

    Definition of the head-stem taper connection. The male stem taper geometry is defined by the proximal and distal diameters and its length. The ‘12/14’ taper is defined with a proximal diameter of 12 mm, a distal diameter of 14 mm and a length of 20 mm, resulting in a male stem taper angle of 5.725° or 5°43’30’’ (the same angle as in the original ‘14/16’ taper). The female taper angle of ceramic heads is always by 1’ to 6’ larger, in order to achieve initial proximal engagement between female and male taper. 6 Most metal heads are designed similarly but some are manufactured with the same nominal female taper angle as the stem. The true contact length depends on the assembly force and the taper angle difference (indicated in green). 85 All these values are not standardized and vary between manufacturers. 86

  • Fig. 2

    Number of peer-reviewed journal publications for the search terms ‘(taper OR trunnion) AND hip AND corrosion’ in PubMed. The number of publications started to increase from the few anecdotal reports published before 2009 and peaked in 2016 with continuing high interest since.

  • Fig. 3

    (a) X-ray of a stem taper fracture in a Ti-alloy primary stem with a large MoM bearing after four years in situ. This design utilizes a Ti-alloy adapter sleeve to assemble the CoCr-bearing surface to the stem. (b) The explanted components with the broken stem taper inside the adapter sleeve inside the head (courtesy H. Ettema, Isala).

  • Fig. 4

    (a) Female head taper corrosion after seven years in situ. The overlay shows the depth of the female head taper wear (material loss 6.7 mm³). (b) The male stem taper did not show signs of corrosion but slight marks from the revision surgery.

  • Fig. 5

    (a) X-ray after fracture of of a Ti-Ti Revision stem after seven years in situ (courtesy M. Rudert, Würzburg). (b) Fracture surfaces of the broken taper showing the typical split lines of a fatigue fracture.

  • Fig. 6

    (a) X-ray of a male patient (76 years) prior the 4th isolated right cup revision in 2019 after original stem implantation with a 32 mm metal XXXL head in 1984. (b) The stem taper showed only light corrosion and minor damage from the multiple revision surgeries. (c) X-ray after the revision with a 36 mm ceramic option XXXL head on the original stem.

  • 1.

    Fricker DC , Shivanath R . Fretting corrosion studies of universal femoral head prostheses and cone taper spigots. Biomaterials 1990 Sep;11:495500 .

  • 2.

    Hussenbocus S , Kosuge D , Solomon LB , Howie DW , Oskouei RH . Head-neck taper corrosion in hip arthroplasty. Biomed Res Int 2015:758123 .

  • 3.

    Piconi C , Streicher RM . Forty years of ceramic-on-ceramic THR bearings. Semin Arthroplasty 2013; 24:188192 .

  • 4.

    Willmann G . Das prinzip der konus-steckverbindung für keramische kugelköpfe bei hüftendendoprothesen. J Mater Sci Technol 1993; 24:315319 .

  • 5.

    Werner PH , Ettema HB , Witt F , Morlock MM , Verheyen CC . Basic principles and uniform terminology for the head-neck junction in hip replacement. Hip Int 2015; 25:115119 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 6.

    National Joint Registry. 7th Annual Report for England and Wales. 2010 .

  • 7.

    McMinn D , Daniel J . History and modern concepts in surface replacement. Proc Inst Mech Eng H 2006; 220:239251. H .

  • 8.

    Garbuz DS , Tanzer M , Greidanus NV , Masri BA , Duncan CP . The John Charnley Award: Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res 2010; 468:318325 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 9.

    Nassutt R , Wimmer MA , Schneider E , Morlock MM . The influence of resting periods on friction in the artificial hip. Clin Orthop Relat Res 2003:127138 .

  • 10.

    Morlock MM , Bishop N , Rüther W , Delling G , Hahn M . Biomechanical, morphological, and histological analysis of early failures in hip resurfacing arthrosplasty. Proc Inst Mech Eng H 2006; 220:333344 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 11.

    Morlock MM , Bishop N & Stahmer F et al. Reasons for failure of hip resurfacing implants. A failure analysis based on 250 revision specimens. Orthopade 2008; 37:695703 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 12.

    National Joint Registry. 6th Annual Report for England and Wales. 2009 .

  • 13.

    Langton DJ , Jameson SS , Joyce TJ , Hallab NJ , Natu S , Nargol AV . Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear. J Bone Joint Surg Br 2010; 92:3846 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 14.

    Langton DJ , Jameson SS & Joyce TJ et al. Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br 2011; 93:10111016 .

  • 15.

    Pandit H , Vlychou M & Whitwell D et al. Necrotic granulomatous pseudotumours in bilateral resurfacing hip arthoplasties: evidence for a type IV immune response. Virchows Arch 2008; 453:529534 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 16.

    Pandit H , Glyn-Jones S & McLardy-Smith P et al. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br 2008; 90:847851 .

  • 17.

    Crawford R , Ranawat CS , Rothman RH . Metal on metal: is it worth the risk? J Arthroplasty 2010; 25:12 .

  • 18.

    Hothi HS , Berber R , Whittaker RK , Blunn GW , Skinner JA , Hart AJ . The relationship between cobalt/chromium ratios and the high prevalence of head-stem junction corrosion in metal-on-metal total hip arthroplasty. J Arthroplasty 2016; 31:11231127 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 19.

    Matthies AK , Racasan R & Bills P et al. Material loss at the taper junction of retrieved large head metal-on-metal total hip replacements. J Orthop Res 2013; 31:16771685 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 20.

    Whittaker RK , Zaghloul AM & Hothi HS et al. Clinical cold welding of the modular total hip arthroplasty prosthesis. J Arthroplasty 2017; 32:610615 .

  • 21.

    Hannemann F , Hartmann A & Schmitt J et al. European multidisciplinary consensus statement on the use and monitoring of metal-on-metal bearings for total hip replacement and hip resurfacing. Orthop Traumatol Surg Res 2013; 99:263271 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 22.

    Australian Orthopedic Association. Annual Report Hip and Knee Arthroplasty of the National Joint Registry 2019. aoanjrr.dmac.adelaide.edu.au/annual-reports-2019; 2019 (date last accessed 28 September 2020).

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 23.

    Australian Orthopaedic Association National Joint Replacement Registry. Metal/Metal Bearing Surface in Total Conventional Hip Arthroplasty 2019. aoanjrr.dmac.adelaide.edu.au/annual-reports-2019; 2019 (date last accessed 28 September 2020).

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 24.

    Cooper HJ , Della Valle CJ & Berger RA et al. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am 2012; 94:16551661 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 25.

    McGrory BJ , MacKenzie J , Babikian G . A high prevalence of corrosion at the head-neck taper with contemporary zimmer non-cemented femoral hip components. J Arthroplasty 2015; 30:12651268 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 26.

    Bansal T , Aggarwal S , Dhillon MS , Patel S . Gross trunnion failure in metal on polyethylene total hip arthroplasty–a systematic review of literature. Int Orthop 2020; 44:609621 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 27.

    Morlock MM , Dickinson EC , Günther KP , Bünte D , Polster V . Head taper corrosion causing head bottoming out and consecutive gross stem taper failure in total hip arthroplasty. J Arthroplasty 2018; 33:35813590 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 28.

    Fahey EJ , Pomeroy E , Rowan FE . Acute trunnion failure of a TMZF alloy stem with large diameter femoral heads. J Orthop 2020; 20:1720 .

  • 29.

    Matsen Ko L , Chen AF , Deirmengian GK , Hozack WJ , Sharkey PF . Catastrophic femoral head-stem trunnion dissociation secondary to corrosion. J Bone Joint Surg Am 2016; 98:14001404 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 30.

    Banerjee S , Cherian JJ & Bono JV et al. Gross trunnion failure after primary total hip arthroplasty. J Arthroplasty 2015; 30:641648 .

  • 31.

    Walker PS , Campbell PA , Brazil D , McTighe T . Trunnion corrosion and early failure in monolithic metal-on-polyethylene tmzf femoral components: A case series. Reconstr Rev 2016; 6:1318 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 32.

    Goldberg JR , Gilbert JL , Jacobs JJ , Bauer TW , Paprosky W , Leurgans S . A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res 2002:149161 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 33.

    Gilbert JL , Buckley CA , Jacobs JJ . In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res 1993; 27:15331544 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 34.

    Lutzner J , Gunther KP , Postler A , Morlock M . Metal ion release after hip and knee arthroplasty - causes, biological effects and diagnostics. Z Orthop Unfall 2020 Aug;158:369382 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 35.

    Panagiotidou A , Meswania J , Hua J , Muirhead-Allwood S , Hart A , Blunn G . Enhanced wear and corrosion in modular tapers in total hip replacement is associated with the contact area and surface topography. J Orthop Res 2013; 31:20322039 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 36.

    Whittaker RK , Hothi HS , Eskelinen A , Blunn GW , Skinner JA , Hart AJ . Variation in taper surface roughness for a single design effects the wear rate in total hip arthroplasty. J Orthop Res 2017; 35:17841792 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 37.

    Hothi HS , Whittaker RK , Meswania JM , Blunn GW , Skinner JA , Hart AJ . Influence of stem type on material loss at the metal-on-metal pinnacle taper junction. Proc Inst Mech Eng H 2015; 229:9197 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 38.

    Porter DA , Urban RM , Jacobs JJ , Gilbert JL , Rodriguez JA , Cooper HJ . Modern trunnions are more flexible: a mechanical analysis of THA taper designs. Clin Orthop Relat Res 2014; 472:39633970 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 39.

    Tan SC , Teeter MG , Del Balso C , Howard JL , Lanting BA . Effect of taper design on trunnionosis in metal on polyethylene total hip arthroplasty. J Arthroplasty 2015; 30:12691272 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 40.

    Norman TL , Denen JE , Land AJ , Kienitz DM , Fehring TA . Taper-trunnion interface stress varies significantly with head size and activity. J Arthroplasty 2019; 34:157162 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 41.

    Del Balso C , Teeter MG , Tan SC , Lanting BA , Howard JL . Taperosis: does head length affect fretting and corrosion in total hip arthroplasty? Bone Joint J 2015; 97-B:911-916 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 42.

    Del Balso C , Teeter MG , Tan SC , Howard JL , Lanting BA . Trunnionosis: does head size affect fretting and corrosion in total hip arthroplasty? J Arthroplasty 2016; 31:23322336 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 43.

    Panagiotidou A , Meswania J & Osman K et al. The effect of frictional torque and bending moment on corrosion at the taper interface : an in vitro study. Bone Joint J 2015; 97-B:463472 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 44.

    Craig P , Bancroft G , Burton A , Collier S , Shaylor P , Sinha A . Raised levels of metal ions in the blood in patients who have undergone uncemented metal-on-polyethylene Trident-Accolade total hip replacement. Bone Joint J 2014; 96-B:4347 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 45.

    Lavernia CJ , Iacobelli DA , Villa JM , Jones K , Gonzalez JL , Jones WK . Trunnion-head stresses in tha: are big heads trouble? J Arthroplasty 2015; 30:10851088 .

  • 46.

    Hothi HS , Panagiotopoulos AC & Whittaker RK et al. Damage patterns at the head-stem taper junction helps understand the mechanisms of material loss. J Arthroplasty 2017; 32:291295 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 47.

    Hothi HS , Eskelinen AP & Berber R et al. Factors associated with trunnionosis in the metal-on-metal pinnacle hip. J Arthroplasty 2017; 32:286290 .

  • 48.

    Krull A , Morlock MM , Bishop NE . Factors influencing taper failure of modular revision hip stems. Med Eng Phys 2018; 54:6573 .

  • 49.

    Buente D , Huber G , Bishop N , Morlock M . Quantification of material loss from the neck piece taper junctions of a bimodular primary hip prosthesis. A retrieval study from 27 failed Rejuvenate bimodular hip arthroplasties. Bone Joint J 2015; 97-B:13501357 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 50.

    Di Laura A , Hothi HS , Henckel J , Kwon YM , Skinner JA , Hart AJ . Retrieval findings of recalled dual-taper hips. J Bone Joint Surg Am 2018; 100:16611672 .

  • 51.

    Tan SC , Lau AC , Del Balso C , Howard JL , Lanting BA , Teeter MG . Tribocorrosion: Ceramic and oxidized zirconium vs cobalt-chromium heads in total hip arthroplasty. J Arthroplasty 2016; 31:20642071 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 52.

    Kurtz SM , Kocagöz SB & Hanzlik JA et al. Do ceramic femoral heads reduce taper fretting corrosion in hip arthroplasty? A retrieval study. Clin Orthop Relat Res 2013; 471:32703282 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 53.

    White PB , Meftah M , Ranawat AS , Ranawat CS . A comparison of blood metal ions in total hip arthroplasty using metal and ceramic heads. J Arthroplasty 2016; 31:22152220 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 54.

    Matharu GS , Pandit HG , Murray DW , Judge A . Adverse reactions to metal debris occur with all types of hip replacement not just metal-on-metal hips: a retrospective observational study of 3340 revisions for adverse reactions to metal debris from the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man. BMC Musculoskelet Disord 2016; 17:495 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 55.

    Hothi HS , Eskelinen AP & Henckel J et al. Effect of bearing type on taper material loss in hips from 1 manufacturer. J Arthroplasty 2018; 33:15881593 .

  • 56.

    Del Balso C , Teeter MG , Tan SC , Lanting BA , Howard JL . Does the additional articulation in retrieved bipolar hemiarthroplasty implants decrease trunnionosis compared to total hip arthroplasty? J Arthroplasty 2018; 33:268272 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 57.

    Moore BF , Lachiewicz PF . Corrosion and adverse tissue reaction after modular unipolar hip hemiarthroplasty. Arthroplast Today 2017; 3:207210 .

  • 58.

    Whitehouse MR , Endo M , Masri BA . Adverse local tissue reaction associated with a modular hip hemiarthroplasty. Clin Orthop Relat Res 2013; 471:40824086 .

  • 59.

    Khair MM , Nam D , DiCarlo E , Su E . Aseptic lymphocyte dominated vasculitis-associated lesion resulting from trunnion corrosion in a cobalt-chrome unipolar hemiarthroplasty. J Arthroplasty 2013; 28:196.e11196.e14 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 60.

    Di Laura A , Hothi H & Henckel J et al. Retrieval analysis of metal and ceramic femoral heads on a single CoCr stem design. Bone Joint Res 2017; 6:345350 .

  • 61.

    Hampton C , Weitzler L , Baral E , Wright TM , Bostrom MPG . Do oxidized zirconium heads decrease tribocorrosion in total hip arthroplasty? A study of retrieved components. Bone Joint J 2019; 101-B:386389 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 62.

    Danoff JR , Longaray J , Rajaravivarma R , Gopalakrishnan A , Chen AF , Hozack WJ . Impaction force influences taper-trunnion stability in total hip arthroplasty. J Arthroplasty 2018; 33:S270S274 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 63.

    Krull A , Morlock MM , Bishop NE . The influence of contamination and cleaning on the strength of modular head taper fixation in total hip arthroplasty. J Arthroplasty 2017; 32:32003205 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 64.

    Lavernia CJ , Baerga L & Barrack RL et al. The effects of blood and fat on Morse taper disassembly forces. Am J Orthop (Belle Mead NJ) 2009; 38:187190 .

  • 65.

    Wuttke V , Witte H , Kempf K , Oberbach T , Delfosse D . Influence of various types of damage on the fracture strength of ceramic femoral heads. Biomed Tech (Berl) 2011; 56:333339 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 66.

    Morlock MM . The taper disaster—how could it happen? Hip Int 2015; 25:339346 .

  • 67.

    Jauch SY , Huber G , Hoenig E , Baxmann M , Grupp TM , Morlock MM . Influence of material coupling and assembly condition on the magnitude of micromotion at the stem-neck interface of a modular hip endoprosthesis. J Biomech 2011; 44:17471751 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 68.

    Schäfer R , Soltész U , Bernard PF . Friction in hip-joint prostheses and its influence on the fixation of the artificial head. J Mater Sci Mater Med 1998; 9:687690 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 69.

    Jauch SY , Coles LG , Ng LV , Miles AW , Gill HS . Low torque levels can initiate a removal of the passivation layer and cause fretting in modular hip stems. Med Eng Phys 2014; 36:11401146 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 70.

    Haschke H , Jauch-Matt SY , Sellenschloh K , Huber G , Morlock MM . Assembly force and taper angle difference influence the relative motion at the stem-neck interface of bi-modular hip prostheses. Proc Inst Mech Eng H 2016; 230:690699 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 71.

    Rehmer A , Bishop NE , Morlock MM . Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses. Clin Biomech 2012; 27:7783 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 72.

    Pennock AT , Schmidt AH , Bourgeault CA . Morse-type tapers: factors that may influence taper strength during total hip arthroplasty. J Arthroplasty 2002; 17:773778 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 73.

    Witt F , Gührs J , Morlock MM , Bishop NE . Quantification of the contact area at the head-stem taper interface of modular hip prostheses. PLoS One 2015; 10:e0135517 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 74.

    Bishop NE , Waldow F , Morlock MM . Friction moments of large metal-on-metal hip joint bearings and other modern designs. Med Eng Phys 2008; 30:10571064 .

  • 75.

    Jauch SY , Huber G & Sellenschloh K et al. Micromotions at the taper interface between stem and neck adapter of a bimodular hip prosthesis during activities of daily living. J Orthop Res 2013; 31:11651171 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 76.

    Hales C , Carroll M , Fryar C , Ogden C . Prevalence of obesity and severe obesity among adults: United States, 2017–2018. NCHS Data Brief 2020;(No. 360, Feb):18 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 77.

    Donaldson FE , Coburn JC , Siegel KL . Total hip arthroplasty head-neck contact mechanics: a stochastic investigation of key parameters. J Biomech 2014; 47:16341641 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 78.

    American Academy of Orthopaedic Surgeons (AAOS). 2019 Annual Report of the American Joint Replacement Registry (AJRR). Rosemont, IL: 2019 .

  • 79.

    Kubacki GW , Gilbert JL . The effect of the inflammatory species hypochlorous acid on the corrosion and surface damage of Ti-6Al-4V and CoCrMo alloys. J Biomed Mater Res A 2018; 106:31853194 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 80.

    Liu Y , Gilbert JL . The effect of simulated inflammatory conditions and Fenton chemistry on the electrochemistry of CoCrMo alloy. J Biomed Mater Res B Appl Biomater 2018; 106:209220 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 81.

    Jennings JM , Dennis DA , Yang CC . Corrosion of the head-neck junction after total hip arthroplasty. J Am Acad Orthop Surg 2016; 24:349356 .

  • 82.

    Endoprothesenregister Deutschland (EPRD). Jahresbericht 2019. Berlin: 2019 .

  • 83.

    Learmonth ID , Young C , Rorabeck C . The operation of the century: total hip replacement. Lancet 2007; 370:15081519 .

  • 84.

    Osman K , Panagiotidou AP , Khan M , Blunn G , Haddad FS . Corrosion at the head-neck interface of current designs of modular femoral components: essential questions and answers relating to corrosion in modular head-neck junctions. Bone Joint J 2016; 98–B:579584 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 85.

    Witt F , Gührs J , Morlock MM , Bishop NE . Quantification of the contact area at the head-stem taper interface of modular hip prostheses. PLoS One 2015; 10:e0135517 .

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 86.

    Mueller U , Braun S , Schroeder S , Sonntag R , Kretzer JP . Same same but different? 12/14 stem and head tapers in total hip arthroplasty. J Arthroplasty 2017; 32:31913199 .

    • PubMed
    • Search Google Scholar
    • Export Citation